View Actual EPA Method 332 (PDF File) - Columbia Analytical ...
View Actual EPA Method 332 (PDF File) - Columbia Analytical ...
View Actual EPA Method 332 (PDF File) - Columbia Analytical ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>EPA</strong> Document #: <strong>EPA</strong>/600/R-05/049<br />
METHOD <strong>332</strong>.0<br />
DETERMINATION OF PERCHLORATE IN DRINKING WATER BY<br />
ION CHROMATOGRAPHY WITH SUPPRESSED CONDUCTIVITY<br />
AND ELECTROSPRAY IONIZATION MASS SPECTROMETRY<br />
Revision 1.0<br />
March 2005<br />
Elizabeth Hedrick and Thomas Behymer, U.S. <strong>EPA</strong>, Office of Research and Development<br />
Rosanne Slingsby, Dionex Corporation<br />
David Munch, U.S. <strong>EPA</strong>, Office of Ground Water and Drinking Water<br />
NATIONAL EXPOSURE RESEARCH LABORATORY<br />
OFFICE OF RESEARCH AND DEVELOPMENT<br />
UNITED STATES ENVIRONMENTAL PROTECTION AGENCY<br />
CINCINNATI, OHIO 45268<br />
<strong>332</strong>.0-1
METHOD <strong>332</strong>.0<br />
DETERMINATION OF PERCHLORATE IN DRINKING WATER BY ION<br />
CHROMATOGRAPHY WITH SUPPRESSED CONDUCTIVITY AND ELECTROSPRAY<br />
IONIZATION MASS SPECTROMETRY<br />
1. SCOPE AND APPLICATION<br />
1.1 This method is applicable to the identification and quantitation of perchlorate (ClO 4- ) in<br />
raw and finished drinking waters. The approach used is ion chromatography with<br />
suppressed conductivity and electrospray ionization mass spectrometry (IC-ESI/MS).<br />
Chemical Abstract Services<br />
Analyte<br />
Registry Number (CASRN)<br />
Perchlorate 14797-73-0<br />
1.2 The ion chromatographic conditions described in this method may be used with a<br />
tandem mass spectrometer (MS/MS) detector as described in <strong>EPA</strong> <strong>Method</strong> 331.0.<br />
Specifically, the IC operational description (Sect. 10) and quality control requirements<br />
(Sect. 9) of <strong>Method</strong> <strong>332</strong>.0 may be used in combination with the MS/MS operational<br />
description and quality control requirements in <strong>Method</strong> 331.0.<br />
1.3 The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets<br />
the Data Quality Objectives (DQOs) that are developed based upon the intended use of<br />
the method. The Lowest Concentration MRL (LCMRL) is the lowest true concentration<br />
for which a future recovery is predicted to fall, with 99 percent confidence, between 50<br />
and 150 percent. The method development laboratory’s LCMRL for ClO 4- , as defined<br />
in Section 3.13, was 0.10 µg/L using the quantitation ion at m/z 101 (Table 5). The<br />
procedure used to determine LCMRLs is described elsewhere. 1<br />
1.4 Laboratories using this method are not required to determine an LCMRL for this<br />
method, but must determine a single laboratory MRL using the procedure described in<br />
Section 9.2.4.<br />
1.5 Detection limit (DL) is defined as the statistically calculated minimum concentration<br />
that can be measured with 99% confidence that the reported value is greater than zero. 2<br />
The DL is compound dependent and is dependent on sample matrix, fortification<br />
concentration, and instrument performance. Determining the DL in this method is<br />
optional (Sect. 9.2.5). The method development laboratory’s DL for ClO 4- in reagent<br />
water was 0.02 µg/L (Table 5).<br />
1.6 The two predominant ClO 4<br />
-<br />
ions that occur naturally at a ratio of 3.086:1 are 35 Cl 16 O 4- ,<br />
m/z 99, and 37 Cl 16 O 4- , m/z 101, respectively. 3 Due to fewer mass spectral interferences,<br />
the concentration of ClO 4<br />
-<br />
using the m/z 101 ion is reported. The m/z 99/101 area count<br />
<strong>332</strong>.0-2
atio and relative retention time are used for confirmation of ClO 4<br />
-<br />
in samples. An<br />
oxygen-18 ( 18 O) enriched ClO 4<br />
-<br />
internal standard is used to improve accuracy and<br />
ruggedness of the method.<br />
1.7 This method is intended for use by or under the supervision of analysts with prior<br />
experience using ion chromatography and mass spectrometry with electrospray<br />
ionization and interpretation of associated data. This method has been developed for<br />
raw and finished drinking waters; however, with further method development the basic<br />
approach may be suitable for measuring ClO 4<br />
-<br />
in other matrices. For example, sample<br />
preparation, sample clean-up and the identification of possible interferences would<br />
require further study. In addition, the IC-ESI/MS conditions may require optimization.<br />
Finally, precision, accuracy and minimum reporting limits would need to be determined<br />
for the matrices of interest.<br />
2.0 SUMMARY OF METHOD<br />
2.1 This method describes the instrumentation and procedures necessary to identify and<br />
quantify low levels of ClO 4- in drinking waters using IC-ESI/MS. Drinking water<br />
samples are collected using a sterile filtration technique. A small volume of sample is<br />
injected into an ion chromatograph. Using an anion exchange column, ClO 4<br />
-<br />
is<br />
separated from constituent cations and anions in the sample using a potassium<br />
hydroxide mobile phase. Due to the use of a non-volatile mobile phase, the eluate from<br />
the column is passed through a conductivity suppressor to remove the potassium (K + )<br />
ions of the mobile phase and to remove the analyte counter cations prior to the eluate<br />
entering the mass spectrometer. An 18 O-enriched 35 Cl 18 O 4- internal standard (m/z 107) is<br />
used for quantitation to improve accuracy and ruggedness of the method. Identification<br />
is made by verifying the relative retention time of the two predominant ClO 4- ions with<br />
respect to the internal standard. Qualitative confirmation of ClO 4- is made by<br />
confirming that the m/z 99/101 area count ratio is within a specified range. If these<br />
conditions are met, along with passing all other QC requirements defined in Section 9,<br />
then the concentration obtained using the m/z 101 quantitation ion is reported.<br />
3. DEFINITIONS<br />
3.1 ANALYSIS BATCH - A sequence of samples, which are analyzed within a 30 hour<br />
period and include no more than 20 field samples. An Analysis Batch must include all<br />
required QC samples, which do not contribute to the maximum field sample total of 20.<br />
The required QC samples include:<br />
• Laboratory Reagent Blank (LRB)<br />
• Continuing Calibration Checks (CCCs)<br />
• Laboratory Fortified Blank (LFB)<br />
• Laboratory Fortified Sample Matrix (LFSM)<br />
• Either a Laboratory Duplicate (LD) or a Laboratory Fortified Sample Matrix<br />
Duplicate (LFSMD)<br />
<strong>332</strong>.0-3
3.2 CALIBRATION STANDARD (CAL) - A solution prepared from the secondary dilution<br />
standard and internal standard. The CAL solutions are used to calibrate the instrument<br />
response with respect to analyte concentration.<br />
3.3 CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing the<br />
method analyte and internal standard, which is analyzed periodically to verify the<br />
accuracy of the existing calibration for the method analyte.<br />
3.4 DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be<br />
identified, measured and reported with 99% confidence that the analyte concentration is<br />
greater than zero. This a statistical determination (Sect. 9.2.5), and accurate quantitation<br />
is not expected at this concentration. 2<br />
3.5 INTERNAL STANDARD (IS) - A pure compound added to all standard solutions and<br />
field samples in a known amount. It is used to measure the relative response of the<br />
method analyte. The internal standard must be a compound that is not a sample<br />
component.<br />
3.6 LABORATORY DUPLICATES (LDs) - Two sample aliquots (LD1 and LD2), taken in<br />
the laboratory from a single sample bottle, and analyzed separately with identical<br />
procedures. Analyses of LD1 and LD2 indicate precision associated specifically with<br />
the laboratory procedures by removing variation contributed from sample collection,<br />
preservation and storage procedures.<br />
3.7 LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other<br />
blank matrix to which known quantities of the method analyte and internal standard are<br />
added in the laboratory. The LFB is analyzed exactly like a sample, including<br />
preservation procedures, and its purpose is to determine whether the method, inclusive<br />
of sample processing, is in control, and whether the laboratory is capable of making<br />
accurate and precise measurements.<br />
3.8 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a field<br />
sample to which a known quantity of the method analyte and internal standard are<br />
added. The LFSM is processed and analyzed exactly like a sample, and its purpose is to<br />
determine whether the sample matrix contributes bias to the analytical results. The<br />
background concentration of the analyte in the sample matrix must be determined in a<br />
separate aliquot and the measured values in the LFSM corrected for background<br />
concentrations.<br />
3.9 LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A second<br />
aliquot of the field sample used to prepare the LFSM which is fortified and analyzed<br />
identically to the LFSM.<br />
3.10 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX (LFSSM) - An aliquot<br />
of the Laboratory Synthetic Sample Matrix Blank (Sect. 3.12) that is fortified with ClO 4<br />
-<br />
<strong>332</strong>.0-4
and processed like a field sample (Sect. 8). It is used to confirm that there is adequate<br />
chromatographic resolution between sulfate and ClO 4- .<br />
3.11 LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other<br />
blank matrix that is treated exactly as a sample including exposure to all filtration<br />
equipment, storage containers and internal standards. The LRB is used to determine if<br />
the method analyte or other interferences are present in the laboratory environment, the<br />
reagents, or apparatus.<br />
3.12 LABORATORY SYNTHETIC SAMPLE MATRIX BLANK (LSSMB) - A solution of<br />
1,000 mg/L each of chloride, sulfate and carbonate (Cl - , SO 4<br />
2-<br />
and CO 3<br />
2-<br />
) anions that is<br />
processed like a field sample. The LSSMB is a reagent blank that must be analyzed<br />
with each LFSSM.<br />
3.13 LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The<br />
single laboratory LCMRL is the lowest true concentration for which a future recovery is<br />
predicted to fall, with 99 percent confidence, between 50 and 150 percent recovery. 1<br />
3.14 MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by<br />
vendors concerning a chemical’s toxicity, health hazards, physical properties, fire, and<br />
reactivity data including storage, spill, and handling precautions.<br />
3.15 MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be<br />
reported as a quantitated value for a target analyte in a sample following analysis. This<br />
defined concentration can be no lower than the concentration of the lowest calibration<br />
standard for that analyte, and can only be used if acceptable quality control criteria for<br />
the analyte at this concentration are met.<br />
3.16 PRIMARY DILUTION STANDARD SOLUTION (PDS) - A solution containing the<br />
method analyte prepared in the laboratory from stock standard solutions and diluted as<br />
needed to prepare calibration solutions and other needed analyte solutions.<br />
3.17 QUALITY CONTROL SAMPLE (QCS) - A solution containing the method analyte at a<br />
known concentration that is obtained from a source external to the laboratory and<br />
different from the source of calibration standards. It is used to verify that the standard<br />
solution has been properly prepared, and stored to maintain its integrity.<br />
3.18 REAGENT WATER (RW) - Purified water which does not contain any measurable<br />
quantity of the method analyte at or above 1/3 the MRL, or interfering compounds that<br />
would affect the determination of the method analyte.<br />
3.19 SECONDARY DILUTION STANDARD (SDS) - A dilution made from the primary<br />
dilution standard (PDS) that is used to prepare the calibration standards.<br />
<strong>332</strong>.0-5
3.20 SELECTED ION MONITORING (SIM) - A mass spectrometric technique where only<br />
one or a few ions are monitored to improve sensitivity.<br />
3.21 STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing the<br />
method analyte that is prepared in the laboratory using assayed reference materials or<br />
purchased from a reputable commercial source.<br />
4. INTERFERENCES<br />
4.1 <strong>Method</strong> interferences may be caused by contaminants in solvents, reagents (including<br />
reagent water), sample bottles and caps, and other sample processing hardware that lead<br />
to discrete artifacts and/or elevated baselines in the chromatograms. All items such as<br />
these must be routinely demonstrated to be free from interferences (less than 1/3 the<br />
MRL for the target analyte) under the conditions of the analysis by analyzing LRBs as<br />
described in Section 9.3.1. Subtracting blank values from sample results is not<br />
permitted.<br />
NOTE: The use of low or high density polyethylene plastic is recommended in place of<br />
glass when possible. If glassware is used, it should be washed with detergent and tap<br />
water and rinsed thoroughly with reagent water since ClO 4- was found in common lab<br />
detergent during method development.<br />
4.2 In anion chromatography, cations are not retained on the analytical column and, in<br />
theory, pass through in the void volume. The anions are separated by charge, size and<br />
polarizability. As a large polarizable molecule, ClO 4- elutes later than the common<br />
inorganic anions (Cl - , SO 4<br />
2-<br />
, CO 3<br />
2-<br />
, and HCO 3- ). Separation of ClO 4<br />
-<br />
from the matrix<br />
ions combined with the specificity of mass spectrometry has resulted in a method that<br />
minimizes interferences for drinking water matrices. There are, however, the following<br />
known conditions or contaminants that, if present, could result in positive or negative<br />
bias in the reporting of ClO 4- .<br />
4.2.1 Direct Chromatographic Co-elution of Contaminants: At sufficiently high<br />
concentration, direct chromatographic co-elution of a contaminant with ClO 4<br />
-<br />
could result in ionization suppression of one or more of the ions of interest<br />
(m/z 99, 101, and/or 107). Alternatively, the contaminant could have the same<br />
m/z as ClO 4- , or in-source collisionally induced dissociation of a co-eluting<br />
contaminant in the ESI interface could produce a fragment ion with the same m/z<br />
as ClO 4- . Any of these conditions could lead to a positive or negative bias of<br />
ClO 4<br />
-<br />
depending on the affected ion.<br />
If a contaminant is present at a concentration detectable by conductivity, a full<br />
mass scan on a replicate analysis may reveal the presence and m/z of the coeluting<br />
contaminant. Direct chromatographic coelution problems or<br />
concentration dependent coelution problems may be solved by achieving<br />
adequate chromatographic separation. This may be done by modifying the<br />
<strong>332</strong>.0-6
eluent strength or modifying the eluent with organic solvents (if compatible with<br />
the IC column and suppressor), changing the detection systems (e.g., MS/MS),<br />
or selective removal of the interference with sample pretreatment. Sample<br />
dilution will only be beneficial if the coelution is a result of column overloading.<br />
High concentrations of polar anions such as pyrophosphate (P 2 O 7<br />
4-<br />
),<br />
tripolyphosphate (P 3 O 10<br />
5-<br />
) and thio compounds, including aromatic sulfonates,<br />
are potential chromatographic interferants. A 75 mM hydroxide mobile phase<br />
concentration was found to elute the polyphosphates well before ClO 4<br />
-<br />
without<br />
compromising data quality.<br />
4.2.2 Concentration-Dependent Interference by Sulfate (SO 4<br />
2-<br />
): Of the common<br />
anions found in drinking waters (Cl - , SO 4<br />
2-<br />
, CO 3<br />
2-<br />
, HCO 3- ), sulfate can be the<br />
most problematic. Sulfate elutes before ClO 4<br />
-<br />
on most of the anion<br />
chromatography columns currently being used for ClO 4<br />
-<br />
analysis; however, it has<br />
a tendency to elute broadly, tailing into the retention time of ClO 4- . Formation of<br />
H 32 SO 4- (m/z 97) and H 34 SO 4- (m/z 99) are favored in the conductivity suppressor<br />
as the pH of the eluate leaving the suppressor becomes strongly acidic. They are<br />
also formed in the electrospray ionization interface. In general, the result of high<br />
sulfate concentrations was observed to be either (1) an inability to detect the<br />
m/z 99 ion, whereas the m/z 101 ion was still detected, or (2) an area count ratio<br />
(m/z 99/101) that did not meet the QC requirement (Sect. 9.3.5). If either of<br />
these effects are observed, the analyst must evaluate the background counts at<br />
m/z 99 in the half minute before ClO 4<br />
-<br />
elutes. If the background counts are high<br />
(approximately 10-20 times higher than the background counts at m/z 99 in the<br />
first CCC of the Analysis Batch, Sect. 10.4.1), sample dilution or pretreatment to<br />
reduce/remove the sulfate is required to meet the m/z 99/101 area count ratio<br />
requirement for confirmation of ClO 4- (Sect. 9.3.5). As the column ages and the<br />
retention time of ClO 4<br />
-<br />
becomes shorter, the analyst might note that the<br />
m/z 99/101 area count ratio is more severely affected by the presence of high<br />
concentrations of sulfate. Column cleaning or replacement is recommended if<br />
this occurs.<br />
4.2.3 ESI/MS Detector Inlet Fouling: The effect of ESI/MS detector inlet fouling is<br />
deterioration of signal intensity for the three ions monitored in this method<br />
(m/z 99, 101 and 107). The deterioration can be rapid (after the analysis of one<br />
problematic matrix) or it can be gradual. To a large extent, the IS will correct<br />
for gradual and minor loss of signal intensity due to ESI/MS inlet fouling.<br />
However, continued loss of signal intensity may eventually affect sensitivity to<br />
the point that it is no longer possible to detect ClO 4- at the MRL, and/or the QC<br />
criteria for IS area counts will fail (Sect. 9.3.4). Not all mass spectrometers<br />
exhibit this problem to the same extent; however, if the problem is observed to<br />
be gradual and significant over the course of a week, it may be greatly reduced<br />
by using an instrument configuration that bypasses the mass spectrometer until<br />
1.5 to 2 minutes prior to the elution of ClO 4<br />
-<br />
(see Figures 1 and 2). This is<br />
<strong>332</strong>.0-7
ecause the ions that have the greatest potential for ESI/MS detector inlet<br />
fouling elute in the first few minutes after sample injection. For the<br />
instrumentation used to collect the data that is presented in this method,<br />
bypassing the mass spectrometer until just prior to the elution of ClO 4<br />
-<br />
dramatically improved system ruggedness and reduced the need for ESI/MS<br />
detector inlet cleaning.<br />
4.2.4 System Carry-over: Carry-over from one analysis may affect the detection of<br />
ClO 4<br />
-<br />
in a second or subsequent analysis. It can occur when the analysis of a low<br />
concentration sample immediately follows the analysis of a high concentration<br />
sample. Carry-over from one analysis to a subsequent analysis may occur if<br />
using an autosampler or if the injection valve is switched back to the load<br />
position too soon after injection of a sample. If ClO 4<br />
-<br />
carry-over is discovered in<br />
blanks proportional to the concentration of the previously injected standard, the<br />
problem must be corrected prior to further analyses.<br />
4.3 Every effort has been made to address known interferences in this method and to inform<br />
the analyst regarding interpretation of chromatographic and mass spectrometric data to<br />
determine if an interferant is present. There are also mandatory QC requirements that, if<br />
failed, should alert the analyst to the possibility of an interferant. Modifications in<br />
sample pretreatment, chromatography and instrumentation are allowed to overcome<br />
interferences.<br />
NOTE: Although modifications are acceptable, the analyst must demonstrate that the<br />
modifications do not introduce any adverse affects on method performance by repeating<br />
and passing all the QC criteria described in Section 9.2, in addition to meeting all the<br />
ongoing QC requirements. Changes are not permitted in sample collection or<br />
preservation (Sect. 8.1).<br />
4.4 The percent of 18 O enrichment of the internal standard may vary between standard<br />
manufacturers. Poor isotopic enrichment may lead to sample contamination by native<br />
Cl 16 O 4- (m/z 99) in the internal standard. Therefore, it must be demonstrated that the IS<br />
does not contain unlabeled ClO 4<br />
-<br />
at a concentration > 1/3 of the MRL when added at the<br />
appropriate concentration to samples (a concentration of 1 µg/L was used during method<br />
development). This is initially confirmed during the IDC and is monitored in each<br />
Analysis Batch by analysis of the Laboratory Reagent Blank (LRB, Sect. 9.3.1).<br />
5. SAFETY<br />
5.1 The toxicity or carcinogenicity of many of the chemicals used in this method have not<br />
been precisely defined; each chemical should be treated as a potential health hazard, and<br />
exposure to these chemicals should be minimized. Each laboratory is responsible for<br />
maintaining awareness of OSHA regulations regarding safe handling of chemicals used<br />
in this method. 4-6 Each laboratory should maintain a file of applicable MSDSs.<br />
<strong>332</strong>.0-8
5.2 Pure ClO 4<br />
-<br />
salts are classified as oxidizers and the potassium hydroxide used in the<br />
mobile phase is caustic. Pure standard materials and stock standards of these<br />
compounds should be handled with suitable protection to skin and eyes.<br />
6. EQUIPMENT AND SUPPLIES (References to specific brands or catalog numbers are<br />
included for illustration only, and do not imply endorsement of the product).<br />
The analytical equipment consists of an ion chromatograph and a mass spectrometer. Figures 1<br />
and 2 show two configurations of the Dionex IC-ESI/MS system that yielded acceptable results<br />
during method development. Figure 1 and Table 1 show the configuration and operating<br />
conditions used to generate the data presented in this method. Table 2 shows the recommended<br />
operating conditions for Metrohm-Peak/Agilent instrumentation. Other instrumentation and<br />
configurations are acceptable provided the QC requirements of the method are met.<br />
6.1 IC-ESI/MS SYSTEM - An analytical system consisting of a microbore chromatographic<br />
pump, a guard and anion separator column, a six-port injection valve, varying sample<br />
loop sizes (50-200 µL), a conductivity suppressor, a conductivity detector and a data<br />
acquisition and management system that has been interfaced with the ESI/MS.<br />
6.1.1 CHROMATOGRAPHIC PUMP - 2-mm isocratic IC pump capable of precisely<br />
delivering flow rates from 0.01-1.0 mL/min [(Dionex Corporation, Sunnyvale,<br />
CA, Model IP25) or an isocratic, metal free, IC pump capable of precisely<br />
delivering flow rates from 0.01-5.0 mL/min, (Metrohm-Peak Inc., Houston, TX,<br />
Model 818) or equivalent].<br />
6.1.2 ANION TRAP COLUMN - A continuously re-generated, high capacity anion<br />
exchange resin column placed before the eluent generator used to remove anions<br />
in the RW (Dionex IonPac CR-ATC-2 mm, Part No. 060477 or equivalent).<br />
6.1.3 ELUENT GENERATOR - An eluent generator is optional (Dionex Model EG40<br />
with EGC-KOH or equivalent). Preparation of mobile phase from high purity<br />
potassium hydroxide (KOH) is permissible. Frequent preparation from KOH<br />
salt may be necessary to maintain a carbonate-free solution.<br />
6.1.4 CHROMATOGRAPHY OVEN - Temperature controlled chromatography oven.<br />
The chromatography oven contains the 6-port injection valve, the guard and<br />
separator columns, the conductivity suppressor and detector. Temperature<br />
maintained at 30 O C is recommended but not required for this method [(Dionex<br />
Model No. LC30) or (Metrohm Advanced IC Separation Center, Metrohm<br />
Model No. 820, Part Nos. 2.820.0220 and 2.833.0010) or equivalent].<br />
6.1.5 ANION GUARD COLUMN - A guard column packed with the same material as<br />
the separator column. It protects the separator column from particulate matter<br />
and compounds that could foul the exchange sites of the separator column<br />
<strong>332</strong>.0-9
[(Dionex AG16, 2-mm internal diameter (I.D.), Part No. 55379) or (Metrohm<br />
ASUPP-4/5 guard, 4-mm I.D., Metrohm Part No. 6.1006.500) or equivalent].<br />
6.1.6 ANION S<strong>EPA</strong>RATOR COLUMN - A 100-250 mm column packed with a solid<br />
phase specially engineered to achieve separation of the anions of interest<br />
[(Dionex AS16, 2-mm I.D. X 250-mm length, Part No. 55378) or (Metrohm<br />
ASUPP5-100, 4-mm I.D. X 100-mm length, Part No. 6.1006.510) or<br />
equivalent].<br />
6.1.7 CONDUCTIVITY SUPPRESSOR - An electrolytic suppressor operated with an<br />
external source of RW. A chemical conductivity suppressor is acceptable,<br />
although sulfuric acid should not be used as the chemical regenerant due to mass<br />
spectrometric interferences caused by HSO 4<br />
-<br />
at m/z 99 [(Dionex Anion Self<br />
Regenerating Suppressor ASRS-MS, 2-mm, Part No. 63008) or (Metrohm<br />
Advanced IC Separation Center, Metrohm Model No. 820, Part No. 2.820.0220<br />
and 2.833.0010) or equivalent].<br />
6.1.8 CONDUCTIVITY DETECTOR - A flow-through detector with an internal<br />
volume that does not introduce analyte band broadening [(Dionex Conductivity<br />
Detector, Model CD25A) or (Metrohm Advanced IC Conductivity Detector,<br />
Metrohm, Model 819, Part No. 2.819.0010) or equivalent].<br />
6.1.9 SAMPLE LOOPS - 50 to 200 µL size. A 200 µL size was used to generate the<br />
data presented in this method. Smaller or larger injection volumes may be used<br />
as long as the Initial Demonstration of Capability (Sect. 9.2), and all calibrations<br />
and sample analyses are performed using the same injection volume.<br />
6.1.10 DATA SYSTEM - Data management software differs from vendor to vendor<br />
and may be recommended by the supplier of the IC or MS. A system that allows<br />
control of both the IC and MS is recommended [(Dionex Chromeleon<br />
Chromatography Management Software, Version 6.4 MSQ) or (Metrohm ICNet<br />
2.3 data management software and Agilent LCMS Chemstation, Metrohm,<br />
v10.02, Part No. G2710AA) or equivalent].<br />
6.1.11 HELIUM - High purity, compressed gas with a pressure of at least 80 psi to<br />
activate valves, sparge eluent and deliver water to the suppressor.<br />
6.1.12 MASS SPECTROMETER - MS equipped with an ESI interface. Operated in<br />
SIM mode [(Dionex Model MSQ-ELMO, manufactured by Thermo Electron,<br />
San Jose, CA) or (Agilent 1100 Series MSD Quad SL, Part No. G1956B,<br />
manufactured by Agilent Technologies, Wilmington, DE) or equivalent].<br />
6.1.13 NITROGEN - Compressed gas for ESI operation, 80 psi. The purity should be<br />
consistent with the MS manufacturer’s recommendations. Due to the high flow<br />
<strong>332</strong>.0-10
ate (>15 L/min), liquid nitrogen or a nitrogen generator is recommended for<br />
long periods of operation.<br />
NOTE: The following instrumentation, used to generate the data presented in this method,<br />
is recommended but not required.<br />
6.1.14 AUXILIARY PUMP - Pump capable of precisely delivering flow rates from<br />
0.01 - 1.0 mL/min. This pump is used to deliver continuous liquid flow to the<br />
mass spectrometer while the eluate flow from the column is diverted to waste<br />
until 1.5 - 2 minutes prior to the elution ClO 4<br />
-<br />
(Dionex high performance<br />
metering pump, Model No. AXP-MS or equivalent). See Figures 1 and 2 for<br />
placement of the pump.<br />
6.1.15 AUXILIARY SIX-PORT VALVE - Electronic, 6-port, rear-loading valve<br />
(Rheodyne, LLC, Rohnert Park, CA, Part No. 9126-038 or equivalent). This<br />
valve may be placed between the exit of the column and the entrance of the<br />
suppressor, as was done for the data reported in this method (Figure 1), or<br />
alternatively, it may be placed between the conductivity detector and the MS<br />
(Figure 2). In the latter configuration, a 50:50 water:acetonitrile mixture is<br />
mixed with the eluate before it enters the MS using a static mixing tee. The flow<br />
rate to the MS during the time of ClO 4<br />
-<br />
elution in Figure 2 is 0.6 mL/min or<br />
0.3 mL/min in Figure 1. As long as all the QC requirements of the method are<br />
met (Sect. 9.2), either configuration is acceptable.<br />
6.1.16 STATIC MIXING TEE - High pressure, microbore, mixing tee. The static<br />
mixing tee is only used in the Figure 2 configuration (UpChurch Scientific, Oak<br />
Harbor, WA, Part No. U466 or equivalent).<br />
6.1.17 AUTOSAMPLER - Used to automate sample analysis. Minimally, the<br />
autosampler should be capable of delivering a volume of sample 10 times the<br />
chosen sample loop size [(Dionex, Model AS40) or (Metrohm Advanced<br />
Sample Processor, Metrohm, Model 788, Part No. 2.788.0010) or equivalent].<br />
6.2 ANALYTICAL BALANCE - Balance capable of +0.1 mg accuracy (Mettler-Toledo,<br />
Inc., Columbus, OH, Mettler AT200 or equivalent).<br />
6.3 STORAGE BOTTLES - Opaque high density polyethylene (HDPE), 30 mL, 125 mL<br />
and 250 mL sizes for storage of standards (Fisher Scientific, Suwanee, GA, Cat. No.<br />
2911974, 2911958 and 2911961 or equivalent).<br />
6.4 SAMPLE CONTAINERS - 125-mL sterile high-density polyethylene (HDPE) bottles<br />
(IChem 125-mL sterile HDPE bottle, Fisher Scientific, Suwanee, GA, Cat. No. N411-<br />
0125 or equivalent) or disposable single-use, sterile polystyrene, 150 mL, with screwcap<br />
for sterile filtered samples (Fisher Scientific, Suwanee, GA, Part No. 09-761-140 or<br />
<strong>332</strong>.0-11
equivalent). The latter can be used directly with a sterile vacuum filter unit if not using<br />
syringe filtration.<br />
6.5 SAMPLE FILTERS - Sterile, single-use, disposable surfactant-free cellulose acetate<br />
(SFCA) 26 mm, 0.2 µm syringe filter (Fisher Scientific, Suwanee, GA, Corning Brand,<br />
Part No. 09-754-13 or equivalent). For samples high in particulates, filters with built-in<br />
prefilters are available. All samples must be filtered at the time of sample collection.<br />
6.6 SYRINGES - Sterile, single-use, disposable, silicone-free, luer-lok, 20 mL (Fisher<br />
Scientific, Suwanee, GA, Target Brand, Part No. 03-377-30 or equivalent).<br />
6.7 SAMPLE PRETREATMENT CARTRIDGES - Single-use, disposable OnGuard-II H<br />
cartridges (Dionex, Part No. 057085 or equivalent) used to remove high concentrations<br />
of carbonate if it is determined to be an interferant. OnGuard-II Ba 2+ cartridges<br />
(Dionex, Part No. 57093 or equivalent) used to remove high concentrations of sulfate if<br />
it is determined to be an interferant. OnGuard-II Ag cartridges (Dionex, Part No. 57089<br />
or equivalent) used to remove high concentrations of chloride if it is determined to be an<br />
interferant. The Ba 2+ pretreatment cartridge is the only one that may be required to meet<br />
the QC requirements of this method (Sect.11.6.2)<br />
6.8 MICRO-PIPETTES - 250 µL, 1000 µL and 10 mL sizes with single-use disposable tips<br />
(Rainin, Oakland, CA, Part Nos. EP-250, EP-1000, and EP-10 mL or equivalent).<br />
6.9 VIALS - Single use, disposable autosampler vials with filter caps, or other disposable,<br />
single use vials with caps having a 10 mL or less capacity to be used for sample<br />
preparation.<br />
7. REAGENTS AND STANDARDS<br />
7.1 REAGENTS AND SOLVENTS - Reagent grade or better chemicals should be used.<br />
Unless otherwise indicated, it is intended that all reagents shall conform to the<br />
specifications of the Committee on <strong>Analytical</strong> Reagents of the American Chemical<br />
Society (ACS), where such specifications are available. Solvents should be HPLC<br />
grade or better. Other grades may be used provided it is first determined that the reagent<br />
is of sufficiently high purity to permit its use without lessening the quality of the<br />
determination.<br />
7.1.1 HIGH PURITY REAGENT WATER (RW) - Purified water which does not<br />
contain any measurable quantity of the target analyte or interfering compounds<br />
at concentrations > 1/3 the MRL for the target analyte. The purity of the water<br />
required for this method cannot be overly emphasized. For this work, deionized<br />
water was further purified using a bench model Millipore water purification<br />
system (Millipore Corp, Billerica, MA, Model No. MilliQ Gradient A10 or<br />
equivalent).<br />
<strong>332</strong>.0-12
7.1.2 ACETONITRILE - ACN, CASRN 75-05-8 (Fisher Scientific, Suwanee, GA,<br />
Cat. No. A998-1 or equivalent). ACN is only required if using the IC-ESI/MS<br />
configuration presented in Figure 2.<br />
7.1.3 METHANOL - MeOH, CASRN 67-56-1 (Fisher Scientific, Suwanee, GA, Cat.<br />
No. A452-1 or equivalent). MeOH is only required if using the Metrohm-Peak<br />
instrumentation.<br />
7.1.4 POTASSIUM HYDROXIDE ELUENT - 75 mM (KOH, F.W.= 56.11, CASRN<br />
1310-58-3, 45% (w/w), Certified ACS Grade , or better). 75 mM KOH is<br />
prepared by diluting 9.35 g of a 45% (w/w) solution to 1 L with RW. Filter,<br />
degas by sonication, or sparge with helium, and pressurize with helium to<br />
minimize absorption of carbon dioxide from the atmosphere. If using an IC<br />
system equipped with an eluent generator (Sect. 6.1.3), KOH eluent preparation<br />
is not necessary.<br />
If using a Metrohm IC system, the recommended eluent is 30 mM NaOH<br />
(NaOH, F.W.= 40.0, CASRN 1310-73-2, 50% (w/w), Certified ACS Grade, or<br />
better) prepared by diluting 2.4 g of the 50% (w/w) solution to 700 mL of RW.<br />
Add 300 mL of MeOH to bring final volume to 1 L. Degas by sonication, or<br />
sparge with helium, to minimize absorption of carbon dioxide from the<br />
atmosphere.<br />
7.1.5 SODIUM SULFATE - Na 2 SO 4 , F.W.=142.04, CASRN 7757-82-6 (Fisher<br />
Scientific, Suwanee, GA., Cat. No. S421-500 or equivalent).<br />
7.1.6 SODIUM CHLORIDE - NaCl, F.W.=58.44, CASRN 7647-14-5 (Fisher<br />
Scientific, Suwanee, GA., Cat. No. S271-500 or equivalent).<br />
7.1.7 SODIUM CARBONATE - Na 2 CO 3 , F.W.=106, CASRN 497-19-8 (Sigma<br />
Aldrich Chemical, St Louis, MO, Cat. No. S6139 or equivalent).<br />
7.2 STANDARD SOLUTIONS - Standard solutions may be prepared from certified,<br />
commercially available solutions or from neat compounds. When a compound purity is<br />
assayed to be 96% or greater, the weight can be used without correction to calculate the<br />
concentration of the stock standard. Solution concentrations listed in this section were<br />
used during the development of this method and are included as an example. Unless<br />
otherwise noted, all standards should be stored in 125-mL HDPE screw-cap bottles<br />
(Sect. 6.3) at 6 O C or less when not in use. Even though stability times for standard<br />
solutions are suggested in the following sections, laboratories should use standard QC<br />
practices to determine when their standards need to be replaced.<br />
7.2.1 INTERNAL STANDARD STOCK STANDARD SOLUTION (IS-SSS) -<br />
1,000 mg Cl 18 O 4- /L. (NaCl 18 O 4 , F.W.=130.4, CASRN 7601-89-0, 90% enriched<br />
on 18 O, 98% pure NaCl 18 O 4 , or better, Isotec, Inc., Miamisburg, OH or<br />
<strong>332</strong>.0-13
equivalent,). A 1,000 mg/L solution of Cl 18 O 4<br />
-<br />
is prepared by dissolving<br />
0.0123 g NaCl 18 O 4 in 10 mL of RW. The solution may be stored in an HDPE<br />
screw-cap bottle (Sect. 6.3). The anhydrous NaCl 18 O 4 salt should be stored in a<br />
desiccator to minimize absorption of water from the atmosphere. The<br />
recommended holding time is one year.<br />
7.2.1.1 INTERNAL STANDARD PRIMARY DILUTION STANDARD -<br />
(IS–PDS) - 1.0 mg Cl 18 O 4- /L. Prepared gravimetrically, using an<br />
analytical balance having +0.1 mg accuracy, by adding 0.1 g (100 µL) of<br />
the IS-SSS to 99.9 g of RW in a 125-mL HDPE storage bottle.<br />
Alternatively, this dilution may be done volumetrically. The<br />
recommended holding time is one year.<br />
7.2.1.2 INTERNAL STANDARD FORTIFICATION SOLUTION - (IS-FS) -<br />
100 µg Cl 18 O 4- /L. Prepared by adding 10 mL of the IS-PDS to 90 mL of<br />
RW in a 125-mL HDPE storage bottle. Alternatively, this dilution may<br />
be done by weight using an analytical balance having +0.1 mg accuracy.<br />
The recommended holding time is one year.<br />
NOTE: A commercially prepared internal standard solution may be<br />
used. (Dionex Corporation, Part No. 062923 or equivalent).<br />
7.2.2 PERCHLORATE STOCK STANDARD SOLUTION (SSS) - 1,000 mg ClO 4- /L.<br />
(NaClO 4 , anhydrous, 99% pure grade, or better, F.W.= 122.4, CASRN<br />
7601-89-0, Sigma Aldrich Co., St. Louis, MO, Cat. No. S-1513, or equivalent).<br />
A 1,000 mg/L solution of ClO 4<br />
-<br />
is prepared by dissolving 0.1231 g of NaClO 4 in<br />
100 mL of RW. The solution may be stored in a HDPE screw-cap bottle<br />
(Sect. 6.3). The anhydrous NaClO 4 salt should be stored in a desiccator to<br />
minimize absorption of water from the atmosphere. The recommended holding<br />
time is one year.<br />
7.2.2.1 PERCHLORATE PRIMARY DILUTION STANDARD - (PDS) -<br />
1.0 mg ClO 4- /L. Prepared gravimetrically using an analytical balance<br />
having +0.1 mg accuracy, by adding 0.1 g (100 µL) of the SSS to 99.9 g<br />
of RW in a 125-mL HDPE storage bottle. Alternatively, this dilution<br />
may be done volumetrically. The recommended holding time is one<br />
year.<br />
7.2.2.2 PERCHLORATE FORTIFICATION SOLUTION - (FS) - 100 µg ClO 4- /L.<br />
Prepared by adding 10 mL of the PDS to 90 mL of RW. The solution may<br />
be stored in a 125-mL HDPE storage bottle. Alternatively, this dilution may<br />
be done by weight using an analytical balance having +0.1 mg accuracy.<br />
The recommended holding time is one year.<br />
<strong>332</strong>.0-14
7.2.3 CALIBRATION STANDARDS (CAL) - The following guide may be used for<br />
preparing 100-mL CAL solutions containing 1.0 µg/L of the IS. The holding<br />
time for CAL solutions is one month.<br />
Final CAL Conc. (µg/L) Volume (mL) FS to add Volume (mL) IS-FS to add<br />
(LRB) 0 0 1.0<br />
0.1 0.1 1.0<br />
0.2 0.2 1.0<br />
0.5 0.5 1.0<br />
1 1 1.0<br />
5 5 1.0<br />
7 7 1.0<br />
10 10 1.0<br />
7.2.4 LABORATORY SYNTHETIC SAMPLE MATRIX (LSSM) - 1,000 mg/L each<br />
of Cl - , SO 4<br />
2-<br />
, CO 3<br />
2-<br />
. Add 1.48 g of Na 2 SO 4 (Sect. 7.1.5), 1.65 g of NaCl<br />
(Sect. 7.1.6) and 1.77 g of Na 2 CO 3 (Sect. 7.1.7) to 1-L volumetric flask and<br />
dilute to volume with RW. The recommended holding time is one year.<br />
7.2.5 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX (LFSSM) -<br />
Prepare an LFSSM at the mid-level concentration of the calibration curve using<br />
the LSSM (Sect 7.2.4). The LFSSM must contain the IS at the same<br />
concentration as the CAL standards. The holding time is one month.<br />
8. SAMPLE COLLECTION, PRESERVATION AND STORAGE<br />
8.1 SAMPLE COLLECTION<br />
8.1.1 Grab samples must be collected in accordance with conventional sampling<br />
practices. 7<br />
8.1.2 When sampling from a cold water tap, open the tap and allow the system to flush<br />
until the water temperature has stabilized (usually approximately 3 to 5<br />
minutes). Collect a representative sample from the flowing system using a<br />
beaker of appropriate size. Use this bulk sample to generate individual samples<br />
as needed. A volume of at least 20 mL is required for each individual sample.<br />
<strong>332</strong>.0-15
8.1.3 When sampling from an open body of water, fill a beaker with water sampled<br />
from a representative area. Use this bulk sample to generate individual samples<br />
as needed. A volume of at least 20 mL is required for each individual sample.<br />
8.1.4 Once representative samples are obtained, they must be filtered to remove any<br />
native microorganisms. Perchlorate is known to be susceptible to microbial<br />
degradation by anaerobic bacteria. 8 Samples are filtered to remove microbes and<br />
stored with headspace to minimize the possibility that anaerobic conditions<br />
develop during storage. At a minimum, leave the top one third of the sample<br />
bottle empty.<br />
8.1.4.1 Remove a sample syringe (Sect. 6.6) from its package and draw up 20<br />
mL of the bulk sample. Remove a sterile sample filter (Sect. 6.5) from<br />
its package without touching the exit Luer connection. Connect the filter<br />
to the syringe making sure that no water from the syringe drops on the<br />
exterior of the filter. For samples high in particulates, pre-filtration using<br />
a sterile filter (0.45 - 10 µm) may help to prevent clogging or rupture of<br />
the 0.2 µm filter. Open a sterile sample container (Sect. 6.4) without<br />
touching the interior. Using gentle pressure, pass the sample through the<br />
filter into the sample container, directing the first milliliter of sample to<br />
waste. During this process do not let the syringe or filter make contact<br />
with the sample container. Following filtration, seal the sample<br />
container tightly, label and prepare the container for shipment. Syringes<br />
and filters are single use items and must be discarded after each sample.<br />
8.2 SAMPLE SHIPMENT AND STORAGE - Samples must be chilled during shipment<br />
and must not exceed 10 C during the first 48 hours after collection. Samples should be<br />
confirmed to be at or below 10 C when they are received at the laboratory. Samples<br />
stored in the lab must be held at or below 6 C until analysis, but should not be frozen.<br />
8.3 SAMPLE HOLDING TIMES - Samples should be analyzed as soon as possible.<br />
Samples that are collected and stored as described in Sections 8.1 and 8.2 may be held<br />
for a maximum of 28 days.<br />
9. QUALITY CONTROL<br />
9.1 QC requirements include the Initial Demonstration of Capability and ongoing QC<br />
requirements that must be met when preparing and analyzing field samples. This<br />
section describes each QC parameter, their required frequency, and the performance<br />
criteria that must be met in order to meet <strong>EPA</strong> quality objectives. The QC criteria<br />
discussed in the following sections are summarized in Section 17, Tables 7 and 8.<br />
These QC requirements are considered the minimum acceptable QC criteria.<br />
Laboratories are encouraged to institute additional QC practices to meet their specific<br />
needs.<br />
<strong>332</strong>.0-16
9.1.1 METHOD MODIFICATIONS - The analyst is permitted to modify IC columns,<br />
mobile phases, chromatographic and ESI/MS conditions. Each time such<br />
method modifications are made, the analyst must repeat the IDC procedures in<br />
Section 9.2.<br />
9.2 INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be<br />
successfully performed prior to analyzing any field samples. Prior to conducting the<br />
IDC, the analyst must first meet the calibration requirements of Section 10.<br />
Requirements for the initial demonstration of laboratory capability are described in the<br />
following sections and are summarized in Table 7.<br />
9.2.1 DEMONSTRATION OF LOW SYSTEM BACKGROUND - Before any<br />
samples are analyzed, or at any time that new reagents, labware or<br />
instrumentation are used, it must be demonstrated that laboratory reagent blanks<br />
are reasonably free of any contaminants that would prevent the determination of<br />
ClO 4- and that the criteria of Section 9.3.1 are met. The LRB and LSSMB must<br />
be filtered using the same sample collection devices that are used for field<br />
samples (Sect. 8.1.4.1).<br />
9.2.1.1 Concentration dependent carry-over is manifest by signals in samples<br />
that increase proportionally to the concentration of the previously<br />
injected sample. Analysis of a blank RW sample must be performed<br />
after the highest CAL standard to assess if carry-over has occurred. This<br />
type of blank is not the same as an LRB in that it is not filtered or<br />
processed as a sample. If there is system carry-over, the source can often<br />
be traced to the use of an autosampler, injection valve problems or an<br />
excess of tubing between the IC and/or MS components. The results for<br />
this sample must meet the criteria outlined in Section 9.3.1. System<br />
carry-over should be eliminated, to the extent possible, by determining<br />
the source of the problem and taking corrective action.<br />
9.2.2 DEMONSTRATION OF PRECISION - Prepare and analyze 7 replicate LFBs<br />
and 7 replicate LFSSMs fortified near the midrange of the Initial Calibration<br />
curve. All samples must be fortified and processed using the sample collection<br />
devices described in Section 8.1.4.1. The relative standard deviation (RSD) of<br />
the measured concentrations at m/z 101 of the replicate analyses must be < 20<br />
percent for both LFB and LFSSM. Calculate %RSD using the equation below:<br />
%RSD = standard deviation of measured concentrations X 100<br />
average measured concentration<br />
<strong>332</strong>.0-17
9.2.3 DEMONSTRATION OF ACCURACY - Using the same set of replicate data<br />
generated for Section 9.2.2, calculate the average percent recovery. The average<br />
recovery must be within 80-120% for both the LFB and the LFSSM data.<br />
Calculate percent recovery (%R) using the following equation:<br />
%R = average measured concentration X 100<br />
fortification concentration<br />
9.2.4 MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Select a target<br />
concentration for the MRL based on the intended use of the method. Establish<br />
an Initial Calibration following the procedure outlined in Section 10.3. The<br />
lowest calibration standard used to establish the Initial Calibration in<br />
Section 10.3 (as well as the low-level CCC) must be at or below the<br />
concentration of the target MRL. Establishing the MRL concentration too low<br />
may cause repeated failure of on-going QC requirements. Confirm the targeted<br />
MRL following the procedure outlined below.<br />
9.2.4.1 Prepare and analyze seven replicate LFBs at the target MRL<br />
concentration. All samples must be processed using the sample<br />
collection devices described in Section 8.1.4.1. Calculate the mean<br />
(Mean) and standard deviation for these replicates using the m/z 101 ion.<br />
Determine the Half Range for the prediction interval of results (HR PIR )<br />
using the equation below:<br />
HR PIR = 3.963 X S<br />
where,<br />
S is the standard deviation, and 3.963 is a constant value for seven<br />
replicates.<br />
9.2.4.2 Confirm that the upper and lower limits for the Prediction Interval of<br />
Results (PIR = Mean + HR PIR ) meet the upper and lower recovery limits<br />
as shown below:<br />
The Upper PIR Limit must be < 150% recovery.<br />
Mean + HR PIR X 100 < 150%<br />
Fortified Concentration<br />
The Lower PIR Limit must be > 50% recovery.<br />
Mean - HR PIR X 100 > 50%<br />
Fortified Concentration<br />
<strong>332</strong>.0-18
9.2.4.3 The target MRL is validated if both the Upper and Lower PIR Limits<br />
meet the criteria described above. If these criteria are not met, the MRL<br />
has been set too low and must be determined again at a higher<br />
concentration.<br />
9.2.5 DETECTION LIMIT DETERMINATION (optional) - While DL determination<br />
is not a specific requirement of this method, it may be required by various<br />
regulatory bodies associated with compliance monitoring. It is the<br />
responsibility of the laboratory to determine if DL determination is required<br />
based on the intended use of the data.<br />
Prepare and analyze at least seven replicate LFBs at a concentration estimated to<br />
be near the Detection Limit over at least 3 days using the procedure described in<br />
Section 11. This fortification level may be estimated by selecting a<br />
concentration with a signal of 2-5 times the noise level.<br />
NOTE: If an MRL confirmation data set meets these requirements, a DL may be<br />
calculated from the MRL confirmation data, and no additional analyses are<br />
necessary.<br />
Calculate the DL using the equation:<br />
DL = S * t ( n - 1, 1 - alpha = 0.99)<br />
where,<br />
t (n-1,1-alpha = 0.99) = Student's t for the 99% confidence level with n-1 degrees of<br />
freedom. Student’s t = 3.143 for n = 7.<br />
n = number of replicates.<br />
S = standard deviation of replicate analyses.<br />
NOTE: Do not subtract blank values when performing MRL or DL calculations.<br />
9.3 ONGOING REQUIREMENTS - This section summarizes the ongoing QC criteria that<br />
must be followed when processing and analyzing field samples. Table 8 summarizes<br />
ongoing QC requirements.<br />
9.3.1 LABORATORY REAGENT BLANK (LRB) - An LRB is analyzed during the<br />
IDC and is required with each Analysis Batch (Sect. 3.1) to confirm that<br />
background contaminants are not interfering with the identification or<br />
quantitation of the method analyte. If the LRB produces a peak within the<br />
retention time window of the analyte that would prevent the determination of the<br />
method analyte, determine the source of contamination and eliminate the<br />
interference before processing samples. The LRB must contain the IS at the<br />
same concentration used to fortify all field samples and CAL standards and must<br />
be processed (i.e., sterile filtration) as described in Section 8.1.4.1. Perchlorate<br />
<strong>332</strong>.0-19
or other interferences in the LRB must be < 1/3 the MRL. If this criterion is not<br />
met, then all data must be considered invalid for all samples in the Analysis<br />
Batch.<br />
NOTE: If samples are collected using devices that have not been previously<br />
evaluated by the laboratory, duplicates of the sample collection devices must be<br />
sent with the samples so an LRB (and an LFB) may be processed in the<br />
laboratory.<br />
NOTE: Although quantitative data below the MRL may not be reliably accurate<br />
enough for data reporting, such data is useful in determining the magnitude of a<br />
background interference. Therefore, blank contamination levels may be<br />
estimated by extrapolation when the concentration is below the lowest<br />
calibration standard<br />
9.3.2 CONTINUING CALIBRATION CHECK (CCC) - CCCs are analyzed at the<br />
beginning of each Analysis Batch, after every ten field samples, and at the end of<br />
the Analysis Batch. See Section 10.4 for concentration requirements and<br />
acceptance criteria.<br />
9.3.3 LABORATORY FORTIFIED BLANK (LFB) - An LFB is required with each<br />
Analysis Batch. The fortified concentration of the LFB must be rotated between<br />
low, medium, and high concentrations from batch to batch. The low<br />
concentration LFB must be as near as practical to the MRL. Similarly, the high<br />
concentration LFB should be near the high end of the calibration range<br />
established during the Initial Calibration (Sect. 10.3). Results of LFBs fortified<br />
at concentrations < the MRL must be recovered within 50-150% of the true<br />
value. Results from the analysis at any other concentration must be recovered<br />
within 80-120% of the true value. If the LFB results do not meet these criteria,<br />
then all data must be considered invalid for all field samples in the Analysis<br />
Batch.<br />
NOTE: LFBs must be processed in the same manner as field samples including<br />
all sample preservation and pretreatment requirements (i.e., sterile filtration) as<br />
described in Section 8.1.4.1.<br />
LFB Fortified Concentration Range<br />
LFB Recovery Limits<br />
< MRL 50 - 150%<br />
>MRL to highest calibration standard 80 - 120%<br />
9.3.4 INTERNAL STANDARD (IS) – The analyst must monitor the peak area of the<br />
internal standard in all injections during each Analysis Batch. The IS response<br />
(as indicated by peak area) for any chromatographic run must not deviate by<br />
<strong>332</strong>.0-20
more than ±30 percent from the area counts measured in the first CCC of the<br />
Analysis Batch (Sect. 10.4.1). If the IS area counts do not meet this criterion,<br />
inject a second aliquot of the sample as part of the same or new Analysis Batch<br />
within the holding time of the sample.<br />
9.3.4.1 If the reinjected aliquot produces an acceptable IS response, report<br />
results for that aliquot.<br />
9.3.4.2 If the IS area counts of the reinjected aliquot still do not meet the IS<br />
criterion, check the IS area of the most recent CCC. If the IS criterion is<br />
met in the CCC but not the sample, report the sample results as “suspect<br />
matrix”.<br />
9.3.4.3 If the IS area criterion is not met in both the sample and the CCC,<br />
instrument maintenance, such as cleaning of the MS sample cone, may<br />
be necessary. Once the analyst has re-established proper operating<br />
conditions, the sample, or affected samples, must be reanalyzed provided<br />
that they are still within their holding times.<br />
9.3.5 AREA COUNT RATIO (m/z 99/101) ACCEPTANCE CRITERIA - All CAL<br />
standards, QC samples and field samples must meet the m/z 99/101 area count<br />
ratio requirement for confirmation of ClO 4- . The measured ratio must fall within<br />
+25% (2.31-3.85). Area count ratios that fall outside this range due to sulfate<br />
interference must be diluted and/or pretreated with barium form pretreatment<br />
cartridges to remove the sulfate to a level that allows better integration of the<br />
ClO 4<br />
-<br />
peak at m/z 99 (Sect. 11.6.2), and thus, better m/z 99/101 area count ratios<br />
for confirmation. If a CAL standard, CCC or LFB fails the area count ratio<br />
acceptance criteria, there may be column, suppressor or instrumental problems.<br />
The source of the problem must be identified and corrected before further<br />
analysis of samples.<br />
9.3.6 RELATIVE RETENTION TIME ACCEPTANCE CRITERIA - Since the<br />
Cl 18 O 4<br />
-<br />
IS has the same retention time as naturally occurring ClO 4- , the retention<br />
time ratio of m/z 99/107 and m/z 101/107 in samples must be within 0.98 - 1.02<br />
(+2% of the ideal ratio of 1) for confirmation of ClO 4<br />
-<br />
in a sample. Use the<br />
equation below to determine the relative retention time:<br />
Relative Retention Time = retention time of m/z 99 or m/z 101 ion<br />
retention time of m/z 107 IS ion<br />
9.3.7 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of an<br />
LFSM (Sect. 3.8) is required in each Analysis Batch and is used to determine<br />
that the sample matrix does not adversely affect method accuracy. If a variety of<br />
different sample matrices are analyzed regularly, for example drinking water<br />
<strong>332</strong>.0-21
from groundwater and surface water sources, performance data should be<br />
collected for each source. Over time, LFSM data should be documented for all<br />
routine sample sources for the laboratory.<br />
9.3.7.1 Within each Analysis Batch, a minimum of one field sample is fortified<br />
as an LFSM for every 20 samples analyzed. The LFSM is prepared by<br />
fortifying a sample with an appropriate amount of the FS (Sect. 7.2.2.2).<br />
Select a fortification concentration that is greater than or equal to the<br />
native background concentration, if known. Selecting a duplicate aliquot<br />
of a sample that has already been analyzed aids in the selection of an<br />
appropriate fortification level. If this is not possible, use historical data<br />
and rotate through low, medium and high calibration concentrations<br />
when selecting a fortifying concentration.<br />
9.3.7.2 Calculate the recovery (%R) for the analyte using the following equation:<br />
%R = (A - B) X 100<br />
C<br />
where,<br />
A = measured concentration in fortified sample<br />
B = measured background concentration in an unfortified aliquot of<br />
the same sample<br />
C = fortification concentration<br />
9.3.7.3 Recoveries for LFSM samples should be 80-120%. Greater variability<br />
may be observed when LFSM samples have ClO 4- concentrations < the<br />
MRL. At these concentrations, LFSM sample recovery should be<br />
50-150%. If the accuracy of ClO 4<br />
-<br />
falls outside the designated range, and<br />
the laboratory performance is shown to be in control in the CCCs, the<br />
recovery is judged to be matrix biased. The result for ClO 4<br />
-<br />
in the<br />
unfortified sample should be labeled “suspect matrix” to inform the data<br />
user that the results are suspect due to matrix effects.<br />
NOTE: A high concentration of sulfate is a known interferant that may<br />
cause the sample to fail the m/z 99/101 area count ratio criteria<br />
(Sect. 9.3.5). In that case, the sample must be diluted or pretreated to<br />
reduce/remove the sulfate to an acceptable level. Refer to Section 11.6<br />
for required remedial action.<br />
NOTE: Field samples that have detectable native ClO 4- concentrations<br />
below the MRL that are fortified at concentrations at or near the MRL<br />
should be corrected for the native levels to obtain more accurate results.<br />
This is the only case where background subtraction of results below the<br />
MRL is permitted.<br />
<strong>332</strong>.0-22
9.3.8 LABORATORY DUPLICATE OR LABORATORY FORTIFIED SAMPLE<br />
MATRIX DUPLICATE (LD or LFSMD) - Within each Analysis Batch, a<br />
minimum of one Laboratory Duplicate (LD) or Laboratory Fortified Sample<br />
Matrix Duplicate (LFSMD) must be analyzed. Laboratory Duplicates check the<br />
precision associated with laboratory procedures. If ClO 4<br />
-<br />
is not routinely<br />
observed in field samples, a LFSMD should be analyzed rather than a LD.<br />
9.3.8.1 Calculate the relative percent difference (RPD) for duplicate measured<br />
concentrations (LD1 and LD2) using the equation:<br />
9.3.8.2 If an LFSMD is analyzed instead of a LD, calculate the relative percent<br />
difference (RPD) for duplicate concentrations of the LFSMs (LFSM and<br />
LFSMD) using the equation:<br />
9.3.8.3 The RPD acceptance criteria for LDs and duplicate LFSMs are listed in<br />
the table below. If the RPD is not within the control, but the laboratory<br />
performance is shown to meet the acceptance criteria in the LFB, the<br />
recovery problem is judged to be matrix related. The result for the<br />
unfortified sample is labeled “suspect matrix” to inform the data user that<br />
the results are suspect due to matrix effects.<br />
Concentration Range<br />
RPD Acceptance Criteria<br />
< 2 X MRL 2 X MRL to highest calibration<br />
standard<br />
1/3 the MRL, then the source of the contamination should be identified and corrected.<br />
The LFSSM should meet the criteria set forth in Sect. 9.3.3. If the LFSSM does not<br />
<strong>332</strong>.0-23
meet the QC acceptance criteria for LFB recovery or if the criteria of Sections 9.3.4 -<br />
9.3.6 fail, instrument maintenance such as column or suppressor cleaning is<br />
recommended.<br />
10. CALIBRATION AND STANDARDIZATION<br />
10.1 Demonstration and documentation of acceptable MS mass calibration and an Initial<br />
Calibration are required before any samples are analyzed. Once the Initial Calibration is<br />
successful, CCCs are required at the beginning and end of an Analysis Batch and after<br />
every tenth field sample. Although not required, it is recommended that the Initial<br />
Calibration be repeated and the mass calibration verified when instrument modifications<br />
(column or suppressor replacement) or maintenance (ESI/MS detector inlet cleaning)<br />
are performed.<br />
NOTE: CAL solutions and CCCs are not processed with the sample collection or<br />
pretreatment devices. This step must be omitted for the CALs and CCCs to identify<br />
potential losses associated with the sample filtration, collection or pretreatment devices.<br />
10.2 MASS CALIBRATION AND INSTRUMENT OPTIMIZATION - MS resolution must<br />
be 1 amu or better. It is recommended that the analyst contact the instrument<br />
manufacturer regarding appropriate mass calibration standards. The user should be<br />
aware that many ESI/MS instruments are designed to analyze macromolecules having<br />
large m/z ratios. As a result, many ESI/MS calibration procedures are designed to cover<br />
the full scanning range of the instrument. Since this method uses the lower portion of<br />
the mass range, it may be necessary to use mass calibration compounds of lower m/z<br />
ratios to achieve a better mass calibration for low m/z ions like ClO 4- . For the<br />
instrumentation used during this method development, a sodium iodide solution was<br />
used as a calibration compound. After the mass calibration has been performed, the<br />
analyst must check mass accuracy for ClO 4- by performing a simple experiment.<br />
Prepare a high CAL standard containing equal amounts of ClO 4- and the IS. While the<br />
CAL standard is being infused, scan over the range of 95 - 115 amu and verify that the<br />
ClO 4- peaks are symmetric about m/z 99, 101 and 107. (There will also be peaks at<br />
m/z 103, 105 and 109 from the internal standard ClO 4<br />
-<br />
ions that have varying numbers<br />
of 18 O atoms.) If the peaks are not symmetric about the mass assignments (i.e., 99 ± 0.3,<br />
101 ± 0.3 and 107 ± 0.3), then a new mass calibration of the MS, or other instrument<br />
maintenance according to the manufacturer’s recommendations, should be performed.<br />
10.2.1 OPTIMIZING MS PARAMETERS - MS instruments have a large number of<br />
parameters that may be varied to achieve optimal signal to noise. Due to<br />
differences in MS design, the recommendations of the instrument manufacturer<br />
should be followed when tuning the instrument. MS conditions may be<br />
established by infusing a solution of ClO 4- , at the same flow rate to be used for<br />
sample analysis, while the analyst optimizes the MS parameters. The cone<br />
voltage determined to be optimal for the instrumentation used in this method<br />
may be adjusted for different MS systems, if necessary, to yield the highest<br />
<strong>332</strong>.0-24
counts for ClO 4<br />
-<br />
at m/z 99 while minimizing in-source collisionally induced<br />
dissociation with subsequent formation of ClO 3<br />
-<br />
(m/z 83).<br />
10.2.2 INSTRUMENT CONDITIONS - Suggested operating conditions are listed in<br />
Table 1 for Dionex instrumentation and in Table 2 for Metrohm-Peak<br />
instrumentation. Conditions different from those described may be used if the<br />
QC criteria in Section 9.2 are met. Different conditions include alternate IC<br />
columns, mobile phases and MS conditions.<br />
10.3 INITIAL CALIBRATION - For the data presented in this method, daily calibrations<br />
were performed using the internal standardization calibration technique; however, it is<br />
permissible to perform an Initial Calibration with daily calibration verification using<br />
CCCs as described in Sections 10.4.1 and 10.4.2. Calibrations must be performed using<br />
peak area (dependent variable) versus concentration (independent variable). Peak<br />
height versus concentration is not permitted.<br />
10.3.1 CALIBRATION SOLUTIONS - Prepare a set of at least five CAL standards as<br />
described in Section 7.2.3. The lowest concentration of the calibration standard<br />
must be at or below the MRL, which will depend on system sensitivity and<br />
intended use of the method. The target MRL must be confirmed using the<br />
procedure outlined in Section 9.2.4 after establishing the Initial Calibration.<br />
Field samples must be quantified using a calibration curve that spans the same<br />
concentration range used to collect the IDC data (Sect. 9.2).<br />
10.3.2 Inject 200 µL of each standard into the IC-ESI/MS. Inject a RW blank after the<br />
highest CAL standard to check for carry-over (Sect. 9.2.1.1). Table 4 is<br />
provided to assist in tabulating data for standards and samples. Tabulate the area<br />
counts of m/z 101 and m/z 107, relative retention time ratios of m/z 99/107 and<br />
m/z 101/107, and the m/z 99/101 area count ratio. Evaluate if the m/z 99/101<br />
area count ratio for all the standards are within the acceptance limits of 2.31 -<br />
3.85 (Sect. 9.3.5) and verify that the relative retention time ratios for m/z 99/107<br />
and m/z 101/107 are between 0.98 - 1.02 (Sect. 9.3.6).<br />
NOTE: A different injection volume may be used as long as the data quality<br />
objectives and QC requirements of the method are met and that the same volume<br />
is used for the analysis of samples.<br />
10.3.3 CALIBRATION ACCEPTANCE CRITERIA - Using the data obtained in<br />
Section 10.3.2, perform a regression (e.g., linear, weighted linear, quadratic) of<br />
the m/z 101/107 area count ratio vs. concentration of ClO 4- . To evaluate if the<br />
chosen regression model yields accurate results across the range, reprocess (do<br />
not re-analyze) CAL standards as unknowns and determine the calculated<br />
concentrations. Determine the percent recoveries of the reprocessed CAL<br />
standards based on the known concentrations. Recoveries at ALL the tested<br />
concentrations must be within 80 - 120% for concentrations > the MRL. For<br />
<strong>332</strong>.0-25
concentrations < the MRL, the minimum acceptance criterion is 50 - 150%<br />
recovery. If the recoveries are not within the acceptable ranges, a different<br />
regression model such as a weighted linear, quadratic or weighted quadratic<br />
should be tested. An acceptable calibration has been obtained when recoveries<br />
of reprocessed standards are within the acceptance criteria stated above.<br />
NOTE: For additional verification of the chosen regression model, or if<br />
experiencing problems in meeting QC criteria contained in this method, refer to<br />
Appendix A for instructions on how to statistically verify regression models for<br />
instrument calibration.<br />
10.3.4 INITIAL CALIBRATION VERIFICATION - Analyze a QCS sample<br />
(Sect. 3.17) fortified near the midpoint of the calibration range. The QCS<br />
sample should be from a source different than the source of the calibration<br />
standards. If a second vendor is not available, then a different lot of the standard<br />
should be used. The QCS should be prepared and analyzed just like a CCC.<br />
The calculated amount of ClO 4- must be 80-120% of the certified value. If the<br />
measured analyte concentration does not meet this criterion, check the entire<br />
analytical procedure to locate and correct the problem before analyzing any field<br />
samples. Calculate percent recovery (%R) using the following equation:<br />
%R = measured concentration X 100<br />
certified concentration<br />
10.4 CONTINUING CALIBRATION CHECKS (CCCs) - At the beginning of the Analysis<br />
Batch, the Initial Calibration must be verified by analyzing a mid-level and MRL level<br />
CCC. Throughout an Analysis Batch the calibration is verified after every ten field<br />
samples by the analysis of a CCC that is rotated between low (< MRL), medium (midlevel<br />
calibration concentration) and high concentration (upper calibration<br />
concentration). CCCs are not counted as samples. Analyze CCCs under the same<br />
conditions used during the Initial Calibration.<br />
10.4.1. MID-LEVEL CCC - The first CCC of an Analysis Batch must be at or near the<br />
mid-point of the calibration to verify the Initial Calibration. Acceptance criteria<br />
for the mid-level CCC is 80-120% recovery. The IS area count acceptance<br />
criterion (Sect. 9.3.4) for subsequent samples must be relative to this first<br />
CCC.<br />
NOTE: If the IS response drifts below 50% of the average IS response of the<br />
CAL standards from the Initial Calibration, instrument maintenance or ESI/MS<br />
detector inlet cleaning may be required (Sect. 4.2.3).<br />
10.4.2 MRL CONCENTRATION CCC- A CCC at a concentration that is < the MRL<br />
concentration is performed, following the mid-level CCC, to verify instrument<br />
sensitivity prior to any analyses. The acceptance criteria is 50-150% recovery.<br />
<strong>332</strong>.0-26
11. PROCEDURE<br />
10.4.3 After every tenth field sample and at the end of an Analysis Batch, CCCs must<br />
alternate between low (< MRL), medium (mid-level calibration concentration)<br />
and high concentration (upper calibration concentration). Calculate the<br />
concentration of ClO 4<br />
-<br />
in the CCCs. A CCC fortified at < MRL must calculate<br />
to be 50-150% of the true value. CCCs fortified at all other levels must calculate<br />
to be 80-120%. If the criteria are not met, then all data from the last successful<br />
CCC to the failed CCC must be considered invalid, and remedial action<br />
(Sect. 10.4.4) should be taken. The remedial action may require re-calibration.<br />
Any field samples that have been analyzed since the last acceptable CCC, that<br />
are still within their holding times, should be reanalyzed after calibration has<br />
been restored.<br />
10.4.4 REMEDIAL ACTION - Failure to meet CCC QC performance criteria may<br />
require remedial action. Major maintenance such as cleaning the ion source or<br />
mass analyzer, requires returning to the Initial Calibration (Sect. 10.3).<br />
11.1 Important aspects of this analytical procedure include proper sample collection and<br />
storage (Sect. 8), ensuring that the instrument is properly calibrated (Sect.10) and that<br />
all required QC are met (Sect. 9.2). This section describes the procedures for sample<br />
preparation and analysis.<br />
11.2 IC-ESI/MS START-UP - The IC should be allowed to operate until the conductivity of<br />
the eluate from the conductivity suppressor stabilizes (
11.3.2 Process all LRBs, LFBs, LSSMBs and LFSSMs using the sample collection<br />
devices is Section 8.1.<br />
11.3.3 Prepare the sample for analysis by pipetting 5 mL into an autosampler vial, or<br />
other suitable single use vial. Dilution of the sample may be required if the<br />
sample concentration is suspected to exceed the upper calibration standard. Add<br />
50 µL of the 100 µg/L IS-FS (Sect. 7.2.1.2), cap the vial and invert several times<br />
to mix. If using a commercially available IS solution, calculate the volume<br />
necessary to achieve a 1.0 µg Cl 18 O 4<br />
-<br />
/L final IS concentration in the sample.<br />
NOTE: A 1% dilution error introduced by the addition of the IS is considered<br />
insignificant. It is permissible to use a different IS concentration; however, the<br />
analyst must be aware that ionization suppression of the native ClO 4<br />
-<br />
may occur<br />
if the IS concentration is too high.<br />
11.4 SAMPLE ANALYSIS<br />
11.4.1 Establish optimal operating conditions for the IC-ESI/MS instrumentation to be<br />
used. Operating conditions may vary depending on instrumentation. The<br />
analyst is responsible for determining optimal conditions for their<br />
instrumentation. The configuration of Figure 1 and the operating conditions of<br />
Table 1 were used to generate the data presented in this method.<br />
11.4.2 Establish a valid Initial Calibration following the procedures outlined in<br />
Section 10.3 or confirm that the calibration is still valid by analyzing the<br />
required CCCs as described in Section 10.4.<br />
11.4.3 Inject aliquots of field samples and QC samples under the same instrumental<br />
conditions used for the Initial Calibration (a 200 µL sample size was used in<br />
collection of data for the method). A sample Analysis Batch is presented in<br />
Table 3.<br />
NOTE: If not using an autosampler, use a syringe to withdraw the sample from<br />
the sample vial. Place the injection valve in the Load position and manually<br />
load the sample loop. The loop size must be the same loop size that was used to<br />
calibrate the instrument. Flush the loop with at least three loop volumes of<br />
sample.<br />
11.4.4 At the conclusion of data acquisition, use the same data acquisition method that<br />
was used for the Initial Calibration to identify peaks in the chromatogram. Use<br />
the data acquisition method to determine the relative retention times and<br />
integrate the peak areas of the monitored ions (m/z 99, 101, and 107).<br />
<strong>332</strong>.0-28
11.5 COMPOUND IDENTIFICATION - Identification/confirmation of ClO 4<br />
-<br />
in a sample is<br />
made by detecting ClO 4<br />
-<br />
at m/z 101 and m/z 99 at the retention time of the internal<br />
standard and by passing the QC criteria established for the m/z 99/101 area count ratio.<br />
11.5.1 RELATIVE RETENTION TIME ACCEPTANCE CRITERIA - Since the<br />
Cl 18 O 4<br />
-<br />
IS has the same retention time as naturally occurring ClO 4- , the retention<br />
time ratio of m/z 99/107 and m/z 101/107 in samples must be within 0.98 - 1.02<br />
(+2% of ideal ratio of 1) for confirmation of ClO 4<br />
-<br />
in a sample.<br />
11.5.2 AREA COUNT RATIO (m/z 99/101) ACCEPTANCE CRITERIA - All CAL<br />
standards, QC samples and field samples must meet the m/z 99/101 area count<br />
ratio requirement for confirmation of ClO 4<br />
-<br />
(Sect. 9.3.5). The measured ratio<br />
must fall within +25% (2.31-3.85). If this area count ratio requirement is not<br />
met for a CCC or LFB, then all samples in the Analysis Batch are considered<br />
invalid and must be reanalyzed after reestablishing acceptable instrument<br />
performance. Field samples having m/z 99/101 area count ratios falling outside<br />
this range due to sulfate interference must be diluted and/or pretreated with<br />
barium form pretreatment cartridges to remove/reduce sulfate to a level that<br />
allows better integration of the ClO 4<br />
-<br />
peak at m/z 99. Section 11.6 describes the<br />
required remedial action in the case that (1) a peak is detected at m/z 101 at the<br />
retention time of the IS at concentrations > the MRL but no peak is detected at<br />
m/z 99 due to high sulfate concentration in the sample, or (2) peaks are detected<br />
at both the m/z 101 and m/z 99 ions but the ratio is not within control due to high<br />
sulfate concentration in the sample. In either case, the required remedial action<br />
described in Section 11.6 must be performed.<br />
11.6 REQUIRED REMEDIAL ACTION - If ClO 4- is detected at m/z 101 at concentrations<br />
> the MRL, but the m/z 99/101 area count ratio fails due to background counts at m/z<br />
99, remedial action is required (Sect. 11.6.1 and/or Sect 11.6.2). Sample dilution and/or<br />
pretreatment using the barium form pretreatment cartridge are acceptable means to<br />
reduce the background at m/z 99 due to high concentrations of sulfate. Generally, the<br />
background at m/z 99 is considered high if it is approximately 10-20 times higher than<br />
the background at m/z 99 measured in the first CCC of the Analysis Batch<br />
(Sect. 10.4.1).<br />
11.6.1 SAMPLE DILUTION - If the concentration detected at m/z 101 is at least<br />
2 times the MRL, a 2-fold dilution of a fresh aliquot of sample may be attempted<br />
to lower the background at m/z 99 due to sulfate in the sample. The m/z 99/101<br />
area count ratio must be re-evaluated in the diluted sample for confirmation of<br />
ClO 4- . If the background at m/z 99 still appears high in the diluted sample,<br />
sample pretreatment using the procedure described in Section 11.6.2 must be<br />
attempted.<br />
NOTE: If a sample is diluted, the analyst must be careful not to dilute the<br />
analyte concentration to below the MRL. Add the IS after dilution.<br />
<strong>332</strong>.0-29
11.6.2 SAMPLE PRETREATMENT - If a sample is pretreated using pretreatment<br />
cartridges, an LRB must also be processed in the same manner as the sample. If<br />
all of the cartridges described in Section 6.7 are used in series, the sample flow<br />
path must be arranged as follows: (1) the Ba 2+ cartridge (used to remove<br />
sulfate), (2) the Ag cartridge (used to remove chloride), (3) a 0.2 µm filter to<br />
remove colloidal silver, and (4) the H + cartridge (used to remove carbonate).<br />
NOTE: Some sample matrices may result in an IS area count QC criteria failure<br />
(Sect 9.3.4), peak shape distortion, high background conductivity, or high<br />
background(s) at m/z 99, 101 and/or 107. In these cases, it may be helpful to use<br />
all three forms of the pretreatment cartridges described in Section 6.7. Consult<br />
the manufacturer’s instructions for preparation of the pretreatment cartridges<br />
prior to use with samples. Generally, the procedure requires rinsing each<br />
cartridge with a minimum volume of RW. It has been found that rinsing with<br />
approximately 2 times the recommended volume of water gives better results.<br />
Insufficiently rinsed cartridges often result in random peaks by conductivity<br />
detection. Add the IS to the sample prior to sample pretreatment using the<br />
cartridges.<br />
11.7 EXCEEDING THE CALIBRATION RANGE - The analyst must not extrapolate<br />
beyond the established calibration range. If the calculated ClO 4- concentration in a<br />
sample is greater than the highest CAL standard of the Initial Calibration, a fresh aliquot<br />
of the sample must be diluted, IS added, and the sample re-analyzed. Incorporate the<br />
dilution factor into the final concentration calculation.<br />
12. DATA ANALYSIS AND CALCULATIONS<br />
12.1 Tabulate data using Table 4 as a guide. Compute sample concentration on the m/z 101<br />
quantitation ion using the calibration generated in Section 10.4.<br />
12.2 If the measured concentration of a field sample exceeds the calibration range, a fresh<br />
aliquot of the sample must be diluted and re-analyzed and pass the confirmation criteria.<br />
12.3 When using an autosampler, the analyst may be unaware that samples continued to be<br />
analyzed even after the failure of on-going QC. Therefore, if using an autosampler,<br />
check that all the on-going QC requirements of the method were successful in the<br />
interim of the analyst’s absence. If a CCC failed at any point during an Analysis Batch,<br />
it will be necessary to re-analyze all samples after the last successful CCC.<br />
12.4 Prior to reporting data, the laboratory is responsible for assuring that QC requirements<br />
have been met or that any appropriate qualifier is documented. Report ONLY those<br />
values that fall between the MRL and the highest calibration standard.<br />
<strong>332</strong>.0-30
12.4.1 Calculations must utilize all available digits of precision, but final reported<br />
concentrations should be rounded to an appropriate number of significant figures<br />
(one digit of uncertainty), with not more than three significant figures.<br />
13. METHOD PERFORMANCE<br />
13.1 SUMMARY - Single laboratory precision in drinking waters, as measured by percent<br />
relative standard deviation (%RSD) of replicate analyses (n=7), was < 10% at<br />
concentrations > 0.2 µg/L ClO 4- . Accuracy, as measured by percent recoveries of<br />
fortified drinking water samples and external Quality Control samples, was 90 - 110%<br />
for concentrations > 0.1 µg/L ClO 4- .<br />
Single laboratory precision in fortified synthetic waters containing up to 1,000 mg/L of<br />
each of the common anions (LFSSM), as measured by %RSD of replicate analyses<br />
(n=7), was 0.1 µg/L ClO 4- . Accuracy, as measured by percent<br />
recovery of fortified synthetic high ionic waters containing up to 1,000 mg/L of each of<br />
the common anions (LFSSM), was 80 - 120% for concentrations > 0.1 µg/L ClO 4- .<br />
13.2 Figure 3 shows chromatograms of a 0.1 µg/L calibration standard with retention times for<br />
the ions monitored in this method (m/z 99, 101 and 107). Figure 4 shows chromatograms<br />
of a 1.0 µg/L ClO 4- LFSSM solution containing 1,000 mg/L of chloride, sulfate and<br />
carbonate. Figure 4 also illustrates the effect of a high background at m/z 99 due to HSO 4<br />
-<br />
.<br />
13.3 Table 5 contains single laboratory DL and LCMRL data in RW.<br />
13.4 Table 6 contains precision (%RSD) and recovery (%R) data for ClO 4<br />
-<br />
in various drinking<br />
water and synthetic water samples at low and high fortification concentrations.<br />
14. POLLUTION PREVENTION<br />
14.1 For information about pollution prevention that may be applicable to laboratories and<br />
research institutions, consult "Less is Better: Laboratory Chemical Management for Waste<br />
Reduction," available from the American Chemical Society's Department of Government<br />
Regulations and Science Policy, 1155 16th Street N.W., Washington D.C. 20036, or on-line<br />
at http://www.ups.edu/community/storeroom/Chemical_Wastes/wastearticles.htm, last<br />
verified in March 2005.<br />
15. WASTE MANAGEMENT<br />
15.1 The analytical procedures described in this method generate relatively small amounts of<br />
waste since only small amounts of reagents are used. The matrices of concern are<br />
finished drinking water. However, the Agency requires that laboratory waste<br />
management practices be conducted consistent with all applicable rules and regulations,<br />
and that laboratories protect the air, water, and land by minimizing and controlling all<br />
releases from fume hoods and bench operations. Also, compliance is required with any<br />
<strong>332</strong>.0-31
16. REFERENCES<br />
sewage discharge permits and regulations, particularly the hazardous waste<br />
identification rules and land disposal restrictions. For further information on waste<br />
management, see the publications of the American Chemical Society’s Committee on<br />
Chemical Safety at http://membership.acs.org/c/ccs/publications.htm, last verified in<br />
March 2005. Or see “Laboratory Waste Minimization and Pollution Prevention,”<br />
Copyright © 1996 Battelle Seattle Research Center, which can be found on-line at<br />
http://www.p2pays.org/ref/01/text/00779/index2.htm, last verified in March 2005.<br />
1. Statistical Protocol for the Determination of the Single-Laboratory Lowest Concentration<br />
Minimum Reporting Level (LCMRL) and Validation of the Minimum Reporting Level<br />
(MRL), available at www.epa.gov/OGWDW/methods/sourcalt.html, last verified in March<br />
2005.<br />
2. Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde, “Trace Analyses for<br />
Wastewaters,” Environ. Sci. Technol. 1981, 15, 1426-1435.<br />
3. Lange's Handbook of Chemistry (15th Edition); Dean, J.A., Ed; McGraw-Hill: New York,<br />
NY, 1999.<br />
4. “Carcinogens-Working with Carcinogens,” Publication No. 77-206, Department of Health,<br />
Education, and Welfare, Public Health Service, Center for Disease Control, National<br />
Institute of Occupational Safety and Health, Atlanta, Georgia, August 1977.<br />
5. “Safety In Academic Chemistry Laboratories,” 3 rd Edition, American Chemical Society<br />
Publication, Committee on Chemical Safety, Washington, D.C., 1979.<br />
6. “OSHA Safety and Health Standards, General Industry,” (29CFR1910). Occupational<br />
Safety and Health Administration, OSHA 2206, (Revised, January 1976).<br />
7. ASTM Annual Book of Standards, Part II, Volume 11.01, D3370-82, “Standard Practice for<br />
Sampling Water,” American Society for Testing and Materials, Philadelphia, PA, 1986.<br />
8. Xu, J., Y. Song, B. Min, L. Steinberg, and B.E. Logan, “Microbial degradation of<br />
perchlorate: principles and applications,” Environ. Engin. Sci, 2003, 20(5), 405-422.<br />
ACKNOWLEDGMENTS<br />
The authors would like to gratefully acknowledge Dr. Douglas W. Later, Dr. William C. Schnute and<br />
Robert J. Joyce of Dionex Corporation for their valuable contributions throughout the development of<br />
this method and in coordinating the collection of second laboratory demonstration data. The authors<br />
also acknowledge Jay Gandhi of Metrohm-Peak, Inc., for providing second laboratory demonstration<br />
data for the method.<br />
<strong>332</strong>.0-32
17. TABLES, DIAGRAMS, FLOWCHARTS AND VALIDATION DATA<br />
TABLE 1. DIONEX IC-MS OPERATING CONDITIONS*<br />
Ion Chromatograph<br />
Dionex Corporation, Sunnyvale, CA<br />
Mobile Phase 75 mM KOH, or 65 mM KOH 1<br />
Guard and Separator Columns<br />
Dionex AG16 + AS16, 250 mm X 2 mm<br />
Flow Rate<br />
0.3 mL/min<br />
Conductivity Suppressor and Current ASRS-MS, 75 mA, or 70 mA 1<br />
Column Temperature<br />
30 O C<br />
Auxiliary Pump Flow Rate 2 0.3 mL/min RW, or 50/50 v/v acetonitrile/water 1<br />
Injection Volume 200 µL<br />
Mass Spectrometer<br />
MSQ with Enhanced Low Mass Option (ELMO)<br />
ThermoFinnigan, San Jose, CA<br />
Ion Energy (V) 0.3<br />
Low Mass Resolution 12.7<br />
High Mass Resolution 12.5<br />
Capillary Voltage<br />
-3 kV<br />
Sampling Cone Voltage<br />
-70 V<br />
Probe Temperature 400 o C, or 500 o C 1<br />
Nitrogen pressure<br />
80 psi<br />
Selected Ion Monitoring m/z 99, 101, 107<br />
Mass Scan Range<br />
0.3 amu<br />
Dwell Time per mass 0.75 sec, or 0.3 sec 1<br />
Smoothing/Points/Range<br />
Boxcar/5/6<br />
1<br />
Condition used in IC-MS Configuration 2.<br />
2<br />
Auxiliary pump is used to deliver RW or 50/50 v/v acetonitrile/water to the conductivity suppressor<br />
and the mass spectrometer until 1.5 minutes prior to the elution of ClO 4<br />
-<br />
depending on which<br />
configuration is used (Figures 1 or 2). For the data presented in this method, RW was used in the<br />
auxiliary pump.<br />
*Instrumentation, when specified, does not constitute endorsement. Brand names are included for<br />
illustration only.<br />
<strong>332</strong>.0-33
TABLE 2. METROHM-PEAK IC-MS OPERATING CONDITIONS*<br />
Ion Chromatograph<br />
Metrohm-Peak, Houston, TX<br />
Mobile Phase<br />
30 mM NaOH/30% Methanol<br />
Guard and Separator Columns Metrohm ASUPP4/5 + ASUPP5-100, 100 mm X 4 mm<br />
Flow Rate<br />
0.7 mL/min<br />
Suppressor Regenerant 60 mM nitric acid/10% Methanol, Rinse 10% Methanol<br />
Column Temperature 30 O C<br />
Injection Volume 100 µL<br />
Mass Spectrometer<br />
1100 Series MSD Quad SL, Agilent Technologies, Wilmington, DE<br />
Low Mass Resolution 0.65<br />
Capillary Voltage<br />
-2 kV<br />
Nitrogen Pressure<br />
80 psi<br />
Fragmentor Voltage 150 V<br />
Drying Gas Temperature 320 o C<br />
Drying Gas Flow Rate 9 - 10 L/min<br />
Selected Ion Monitoring m/z 99, 101, 107<br />
Mass Scan Range<br />
0.1 amu<br />
Dwell Time Per Mass 0.25 sec<br />
*Instrumentation, when specified, does not constitute endorsement. Brand names are included for<br />
illustration only.<br />
<strong>332</strong>.0-34
TABLE 3. SAMPLE ANALYSIS BATCH<br />
Injection # Sample Description Acceptance Criteria Remedial Action<br />
1 Mid-Level CCC 80 - 120 % recovery using<br />
Initial Calibration<br />
Instrument maintenance and<br />
recalibration.<br />
2 MRL CCC 50 - 150% recovery Instrument maintenance to<br />
recover sensitivity and<br />
recalibration.<br />
3 LRB MRL to highest CAL std<br />
50 - 150% recovery<br />
80 - 120% recovery<br />
Identify and correct source of<br />
problem.<br />
5<br />
.<br />
.<br />
14<br />
Field Samples 1 - 10<br />
Pass RT, m/z 99/101 area<br />
count ratio, and IS area count<br />
QC criteria at concentrations<br />
>MRL concentration.<br />
If problem is due to sulfate,<br />
clean up sample using Ba form<br />
cartridge, otherwise report.<br />
15 CCC (rotating<br />
concentrations)<br />
16 LFSM of a field sample<br />
previously analyzed<br />
17 Laboratory Duplicate or a<br />
LFSMD of field sample<br />
previously analyzed.<br />
Choose LFSMD if<br />
samples are low in<br />
perchlorate.<br />
80 - 120% recovery using<br />
Initial Calibration for<br />
concentrations > MRL<br />
50-150% recovery for<br />
concentrations < MRL<br />
At fortification concentrations<br />
> MRL concentration, 80-<br />
120% recovery.<br />
At fortification concentrations<br />
< MRL, 50-150% of true<br />
value.<br />
RPD 2 X MRL<br />
RPD MRL<br />
Instrument maintenance and<br />
recalibration.<br />
<strong>332</strong>.0-35
TABLE 4. EXAMPLE TEMPLATE FOR TABULATION OF SAMPLE DATA FOR QC<br />
REQUIREMENTS<br />
Sample<br />
ID<br />
Ion<br />
(m/z)<br />
Area<br />
Counts<br />
Retention<br />
Time (min)<br />
Relative Retention<br />
Time Ratio<br />
(m/z 99/107,<br />
m/z 101/107)*<br />
Area Count Ratio<br />
(m/z 99/101)**<br />
99<br />
101<br />
107 Are area counts of IS +30% of first CCC<br />
99<br />
101<br />
107 Are area counts of IS +30% of first CCC<br />
* - Acceptance Criteria (0.98 - 1.02)<br />
** - Acceptance Criteria (2.31 - 3.85)<br />
<strong>332</strong>.0-36
TABLE 5. DETECTION LIMIT AND LCMRL FOR PERCHLORATE IN REAGENT<br />
WATER<br />
Detection Limit* 0.02<br />
LCMRL 0.10<br />
Concentration<br />
(µg/L) - m/z 101<br />
*Fortification concentration - 0.05 µg/L.<br />
Seven replicates over three days.<br />
TABLE 6. PRECISION AND RECOVERY DATA FOR PERCHLORATE IN VARIOUS<br />
MATRICES (N=7)<br />
Matrix<br />
Background<br />
Conc. (µg/L)<br />
Fortification<br />
Conc. (µg/L)<br />
m/z 99/101<br />
Area Ratio<br />
Avg. %<br />
Recovery<br />
%RSD<br />
Reagent Water ND 1 0.50 3.05 102 3.6<br />
0.05 2.77 105 14<br />
LSSM<br />
ND<br />
0.20 2.64 90 11<br />
1.0 2.70 90 3.0<br />
Surface Source Tap<br />
Water<br />
0.27 1.0 2.98 99 1.6<br />
High TOC Surface<br />
Source Tap Water 2<br />
ND<br />
0.20 2.93 104 8.6<br />
1.0 3.04 95 1.5<br />
Ground Water
TABLE 7. INITIAL DEMONSTRATION OF CAPABILITY (IDC) REQUIREMENTS<br />
<strong>Method</strong><br />
Reference<br />
Requirement Specification and Frequency Acceptance Criteria<br />
Section<br />
9.2.1<br />
Demonstration of Low System<br />
Background<br />
Analyze an LRB and LSSMB prior to any<br />
other IDC steps and when modifications<br />
are made.<br />
TABLE 8. ON-GOING QUALITY CONTROL REQUIREMENTS (SUMMARY)<br />
<strong>Method</strong><br />
Reference<br />
Requirement Specification and Frequency Acceptance Criteria<br />
Section 8<br />
Sample Collection, Preservation, and<br />
Holding Time<br />
28 days, samples must be sterile filtered<br />
through a 0.2 µm filter with the filtrate<br />
collected in a sterile bottle.<br />
Ship at < 10 o C to be received within 48 hours.<br />
Once received at the lab, samples should be<br />
analyzed as soon as possible. Sterile filtered<br />
samples must be stored with head space. Leave<br />
1/3 of bottle empty. Store at 6 o C or less.<br />
Section<br />
10.3<br />
Initial Calibration<br />
Use internal standardization calibration<br />
and a minimum of 5 calibration standards.<br />
Use peak area for calibration and<br />
quantitation.<br />
80 - 120% recovery of all reprocessed standards<br />
at > the MRL.<br />
50 - 150% recovery of reprocessed standards <<br />
the MRL.<br />
Section<br />
9.3.1<br />
Laboratory Reagent Blank (LRB)<br />
Analyze one LFB per Analysis Batch<br />
(every 20 field samples).<br />
Demonstrate that the target analyte is
TABLE 8 (Continued). ON-GOING QUALITY CONTROL REQUIREMENTS (SUMMARY)<br />
<strong>Method</strong><br />
Reference<br />
Requirement Specification and Frequency Acceptance Criteria<br />
Section<br />
9.3.5<br />
Area Count Ratio (m/z 99/101)<br />
Acceptance Criteria<br />
In all standards, field samples, LFBs,<br />
CCCs, etc., analyzed during each<br />
Analysis Batch.<br />
The calculated m/z 99/101 area count ratio<br />
must be within + 25% (2.31 - 3.85).<br />
Section<br />
9.3.6<br />
Relative Retention Time Acceptance<br />
Criteria<br />
In all standards, field samples, CCCs,<br />
LFBs, LFSMs, etc.<br />
Relative retention times of m/z 99/107 and<br />
m/z 101/107 must be within 0.98 - 1.02.<br />
Section<br />
9.3.7<br />
Laboratory Fortified Sample Matrix<br />
(LFSM)<br />
Analyze one LFSM per Analysis Batch<br />
(every 20 field samples) fortified with<br />
perchlorate at a concentration that is<br />
greater than or equal to the native<br />
concentration.<br />
Recoveries not within 80 - 120% of the<br />
fortified amount at > MRL may indicate a<br />
matrix effect. Conc. < MRL may be<br />
recovered 50 - 150%.<br />
Section<br />
9.3.8<br />
Laboratory Duplicate (LD) or<br />
Laboratory Fortified Sample Matrix<br />
Duplicate (LFSMD)<br />
Analyze at least one LFSMD or LD with<br />
each Analysis Batch of up to 20 field<br />
samples.<br />
FIGURE 1. IC-ESI/MS Configuration Used to Generate Data in <strong>Method</strong><br />
FIGURE 2. Alternative IC-ESI/MS Configuration<br />
* static mixing tee<br />
<strong>332</strong>.0-41
FIGURE 3. MASS CHROMATOGRAM OF A STANDARD CONTAINING 0.1 µg/L ClO 4<br />
-<br />
AND 1.0 µg/L INTERNAL STANDARD<br />
<strong>332</strong>.0-42
FIGURE 4. MASS CHROMATOGRAM OF AN LFSSM CONTAINING 1.0 µg/L ClO 4<br />
-<br />
AND<br />
1.0 µg/L INTERNAL STANDARD<br />
<strong>332</strong>.0-43
APPENDIX A (Optional)<br />
Statistical Validation of the Regression Model Used For Instrument Calibration<br />
Introduction<br />
Selection of an appropriate regression model for instrument calibration is critical for obtaining accurate,<br />
non-biased results and for the determination of an LCMRL and MRL. The following guidance is<br />
provided for labs that desire additional validation of their instrument calibration model and for labs that<br />
may be experiencing problems meeting the QC criteria contained in Section 9 of the method.<br />
Background<br />
The Calibration Range (CR) is defined as the concentration range over which the instrument has been<br />
calibrated and results may be reported. During the Initial Demonstration of Capability (Sect. 9.2), the<br />
regression model used to describe the CR must pass certain minimum criteria for accuracy as determined<br />
by the percent recovery of standards that are reprocessed (not re-analyzed) as samples. In practice, a lab<br />
may find that low concentration standards are consistently biased low (or high). It may be that the<br />
analyst has attempted to calibrate over too large of a concentration range for the chosen model. The<br />
analyst may change the range of interest or select a different model but be unsure if the new range and/or<br />
model is appropriate. The F-test for lack of fit, described below, is a statistical metric for determining if<br />
a selected regression model (e.g., linear, quadratic) gives a non-biased estimate of the expected response,<br />
Pred Y (area count ratio, m/z 101/107), as a function of standard concentration, x.<br />
General Recommendations<br />
! The instrument should be operationally stable. This may require a period of approximately<br />
30 minutes of operation with liquid flow through the IC-ESI/MS system.<br />
! If the desired CR is two orders of magnitude or greater, a weighted regression model will<br />
likely be required. Most newer instrumentation automatically allows for this type of<br />
calibration. For the instrumentation used in this method, it was found that a non-weighted<br />
linear regression model yielded 90-110% recoveries of all standards, for concentrations 0.1-<br />
1.0 µg/L ClO 4- . If, however, the upper range was extended to 5.0 or 10.0 µg/L, a weighted<br />
linear regression was required (weight factor = 1/x) to achieve the same results. Using a nonweighted<br />
linear regression across the range 0.1-5.0 or 10.0 µg/L resulted in consistently high<br />
recoveries (115-131%) for the 0.1 µg/L CAL standards. Choosing a short range over which<br />
the variance is constant (e.g. 0.1-1.0 µg/L ClO 4- ), or using a weighted regression model, are<br />
both acceptable means to obtain a regression that yields accurate, non-biased results across the<br />
range.<br />
! The F-test for lack of fit assumes constant variance across the concentration range. Statistical<br />
software is highly recommended to test for constant variance and to perform the F-test for lack<br />
of fit; however, if statistical software is not available, the mathematical procedures described<br />
below should result in selection of an appropriate regression model.<br />
<strong>332</strong>.0-44
Procedure<br />
1. Prepare and inject, in duplicate, five standards that span the range of interest. Concentration is<br />
the independent variable, x, and the dependent variable, y, is the area count ratio m/z 101/107.<br />
Evaluate each standard to make sure the IS area counts are within control and that the m/z<br />
99/101 area count ratio is in control. Tabulate concentration (µg/L), x, and response (area<br />
count ratio m/z 101/107), y.<br />
NOTE: To perform the F-test for lack of fit, there must be replicates on some or all of the<br />
levels of concentration, x.<br />
2. Using the data obtained in Step 1, perform a non-weighted linear regression of the area count<br />
ratio (area count ratio m/z 101/107) vs. concentration (µg/L) of ClO 4- .<br />
3. Decide what will be deemed acceptable recoveries for the data quality objectives of the work.<br />
For the example presented below, it was decided that 90 - 110% recoveries across the range<br />
would be the criteria for accepting the model (see Table A1).<br />
4. To evaluate if the chosen regression model yields accurate results (i.e., constant variance<br />
across the range), reprocess (do not re-analyze) standards as unknowns and determine the<br />
calculated concentrations. Determine the percent recoveries of the reprocessed standards<br />
based on the known concentrations (see example in Table A1). Recoveries should meet the<br />
recovery criteria and be consistent across the range, i.e., recoveries at ALL the tested<br />
concentrations must be within the recovery range (in the example presented here that range is<br />
90 - 110%).<br />
5. If the recoveries are not consistent across the range, a weighted linear regression model should<br />
be tested. Reprocess the data and re-evaluate the recalculated recoveries. If the results are still<br />
unacceptable, delete the highest standard from the regression model and reprocess the data. If<br />
unacceptable results are still encountered when the range has been reduced to one order of<br />
magnitude, there may be very poor precision between duplicate analyses. This may signal that<br />
instrument maintenance is required.<br />
6. Since the F-test for lack of fit assumes normally distributed data with equal variances for the<br />
Y distribution (i.e., across the range of concentrations), a weighted regression model should be<br />
tried before proceeding to the F-test for lack of fit. Weighting will generally give better<br />
recoveries across a wide calibration range. When an acceptable range and model have been<br />
chosen that yields recoveries of reprocessed standards within the set criterion for recoveries of<br />
reprocessed standards, proceed to Step 7, the F-test for lack of fit.<br />
7. F-TEST FOR LACK OF FIT - The use of statistical software to perform the F-test for lack of<br />
fit is highly recommended. If this option is not available, however, use a spreadsheet software<br />
program and the following directions to perform the test. Prepare a table exactly like Table A2<br />
and enter the data required in each column.<br />
<strong>332</strong>.0-45
The test statistic involves calculating F* for the chosen model and comparing it to a critical F<br />
value from a standard table of F values. 1 The test statistic is as follows:<br />
F* = SSLF / DF LOF<br />
SSPE / DF PE<br />
where,<br />
F* = calculated F for regression model<br />
SSLF = lack of fit sum of squares. See Table A2 for calculation.<br />
SSPE = pure error sum of squares. See Table A2 for calculation.<br />
DF LOF = degrees of freedom for SSLF. Equals c - 2 for 1 st order polynomial.<br />
Equals c - 3 for second order polynomial (quadratic).<br />
DF PE = degrees of freedom for SSPE. Equals n - c.<br />
c<br />
n<br />
= number of concentration levels.<br />
= total number of observations.<br />
Table A2 shows the mathematical calculation of SSLF and SSPE from the data obtained from<br />
the chosen regression model (a weighted 1 st order linear regression model). The table was<br />
completed by entering the required data into a software spreadsheet program. In the example<br />
provided in Table A2, n = 10 and c = 5. Critical F(1-alpha, c-2, n-c) = F(0.95, 3, 5) = 9.01.<br />
The Decision rule was:<br />
If F* < 9.01, then conclude that the regression model is appropriate.<br />
If F* > 9.01, then conclude that the regression model is not appropriate.<br />
In this example, the calculated F* using a weighted linear regression model was 0.8874 which<br />
is less than the critical F value of 9.01. It was concluded that the selected model was<br />
appropriate. If the calculated F* had been greater than the critical F value, then a different<br />
model (quadratic or weighted quadratic) would have been evaluated and consistent recoveries<br />
and lack of fit would have been tested again with proper modification of the degrees of<br />
freedom for SSLF. Return to Step 4.<br />
Reference<br />
1. Neter, J., W. Wasserman and M. Kutner, Applied Linear Regression Models, 1989, Irwin, Inc,<br />
Boston, MA.<br />
<strong>332</strong>.0-46
APPENDIX A<br />
TABLE A1. SAMPLE DATA COMPILATION AND DETERMINATION OF ACCURACY OF<br />
CALIBRATION MODEL<br />
A B C<br />
1 X = Conc. 1<br />
µg/L<br />
Pred Xij 2<br />
µg/L<br />
%Recovery<br />
2 0.1 0.1094 109<br />
3 0.1 0.0926 92.6<br />
4 0.5 0.4828 96.6<br />
5 0.5 0.4609 92.2<br />
6 1 1.0196 102<br />
7 1 0.9882 98.8<br />
8 5 5.0781 102<br />
9 5 5.0889 102<br />
10 10 10.155 102<br />
12 10 10.349 104<br />
1<br />
X = concentration of CAL standard. Levels of X = 1 - j, replicates = 1 - i<br />
2<br />
Predicted Xij = concentration calculated from regression model for a given Yij.<br />
NOTE: The weighted linear regression equation was y = 0.0014148 + 0.3612397 X.<br />
<strong>332</strong>.0-47
APPENDIX A<br />
TABLE A2. SPREADSHEET TABULATION OF DATA TO DETERMINE F-TEST FOR LACK OF FIT<br />
1 1 X<br />
A B C D E F G<br />
2 1/Xij Yij=(m/z 101/107)*(1/X)<br />
2 Y<br />
4 Pred Y<br />
3 Mean Yj<br />
4 (Pred Yij)*(1/X) (MeanYj - PredYij)^2 (Yij-MEAN Yj)^2<br />
3 10 0.409412140738561 0.3790897717 0.3753877 1.37053349457528E-005 0.000919446063506114<br />
4 10 0.348767402681377 0.3790897717 0.3753877 1.37053349457528E-005 0.000919446063506114<br />
5 2 0.35171017276265 0.3437878062 0.3640693 0.000411338989721146 6.27638915474167E-005<br />
6 2 0.335865439688544 0.3437878062 0.3640693 0.000411338989721146 6.27638915474176E-005<br />
7 1 0.369735548824696 0.3640767720 0.3626545 2.02285782420544E-006 3.20217544267144E-005<br />
8 1 0.358417995303424 0.3640767720 0.3626545 2.02285782420544E-006 3.20217544267138E-005<br />
9 0.2 0.367171010012092 0.3675596088 0.36152266 3.64447517258711E-005 1.51009076672299E-007<br />
10 0.2 0.367948207738986 0.3675596088 0.36152266 3.64447517258711E-005 1.51009076672299E-007<br />
11 0.1 0.366984493885163 0.3705014438 0.36138118 8.31792126127104E-005 1.23689370239679E-005<br />
12 0.1 0.374018393805966 0.3705014438 0.36138118 8.31792126127104E-005 1.23689370239683E-005<br />
13 5 SSLF = 0.00109338229365937 0.00205350331116177 = 6 SSPE<br />
14 SSLF/c-2 = 0.000364460764553124 0.000410700662232354 =SSPE/n-c<br />
15<br />
7 F* = (SSLF/3)/(SSPE/5)= 0.887<br />
1<br />
X = concentration of CAL standard with weighting factor applied. Levels of X = 1 - j, replicates = 1 - i. NOTE: If not using a weighted regression, then X = concentration.<br />
2<br />
Y = (Yij m/z 101/107area count ratio) times (the weighting factor, 1/X). NOTE: If not using a weighted regression, then Y = m/z 101/107 area count ratio.<br />
3 Mean Yj = mean of Y for a given replicate level, i.<br />
4 Pred Y = predicted Yij using the chosen regression model for a given X with weighting factor applied. NOTE: If not using a weighted regression, then Pred Y would not<br />
have the weighting factor applied. The weighted linear regression equation was y = 0.0014148 + 0.3612397 X.<br />
5 SSLF = lack of fit sum of squares. Obtained by summing cells E3..E12. Degrees of freedom = c - 2 for 1 st order polynomial. For this example, degrees of freedom for SSLF = 3.<br />
NOTE: Degrees of freedom for a quadratic fit would be c - 3.<br />
6 SSPE = sum of squares pure error. Obtained by summing cells F3..F12. For this example, degrees of freedom for SSPE = 5.<br />
7<br />
F* = calculated F for the given regression model.<br />
<strong>332</strong>.0-48